Severity
Moderate
FDA Devices recall · Reported May 20, 2026
Software issue: When programming the SynchroMed Pump with the A810 CP App in Flex Infusion Mode, the infusion steps should automatically be arranged in consecutive order, from ea…
Medtronic Neuromodulation recalled Medtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application,… - a moderate-severity action.
Medtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application,… was recalled by Medtronic Neuromodulation in May 20, 2026. Reason: Software issue: When programming the SynchroMed Pump with the A810 CP App in Flex Infusion Mode, the infusio…. Check the official notice for the remedy. Verify recall #Z-2201-2026 with the FDA Devices before acting.
The recall
Medtronic Neuromodulation issued this moderate-severity FDA Devices recall-Software issue: When programming the SynchroMed Pump with the A810 CP App in Flex Infusion Mode, the infusio….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2201-2026 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2201-2026) was formally reported on May 20, 2026, with the manufacturer initiating the action on April 8, 2026. It is classified under Moderate severity (Class II), with a current status of Ongoing. Medtronic Neuromodulation is listed as the recalling firm, operating out of Minneapolis, MN. Federal records list the affected scope as 10165 units.
The documented reason for this recall is: Software issue: When programming the SynchroMed Pump with the A810 CP App in Flex Infusion Mode, the infusion steps should automatically be arranged in consecutive order, from earliest to latest. However, in rare cases… Distribution data in the federal record shows the product reached: Worldwide distribution.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
10165 units
Related Recalls
6
3 from same agency
Medtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application, loaded on the following devices: 1. PROG CT900A CLINICIAN TABLET US, Model Number CT900A; 2. PROG CT900B CLINICIAN TABLET EU -UK, Model Number CT900B; 3. PROG CT900C CLINICIAN TABLET UK, Model Number CT900C; 4. PROG CT900D CLINICIAN TABLET GLOBAL, Model Number CT900D; 5. PROG CT900E CLINICIAN TABLET GLBL SMSNG, Model Number CT900E; 6. PROG CT900F CLINICIAN TABLET GLBL SMSNG, Model Number CT900F; implanted programmable infusion pump
Software issue: When programming the SynchroMed Pump with the A810 CP App in Flex Infusion Mode, the infusion steps should automatically be arranged in consecutive order, from earliest to latest. However, in rare cases, the steps may be displayed out of order. If this occurs, the programmed Flex Infusion schedule will not be delivered at the intended time, if at all.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-2201-2026 |
| Date reported | May 20, 2026 |
| Date initiated | April 8, 2026 |
| Recalling firm | Medtronic Neuromodulation |
| Firm location | Minneapolis, MN |
| Affected scope | 10165 units |
| Distribution | Worldwide distribution. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported May 20, 2026. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.