Severity
Moderate
FDA Devices recall · Reported June 10, 2020
The top label for dental implant may incorrectly indicate the implant length. Specifically, the label may contain 11.5mm for the length, which is shorter than the actual length of…
Nobel Biocare Usa Llc recalled NobelActive Internal NP 3.5x15mm Endosseus dental implant, REF: 34128, Rx Only, Sterile R… - a moderate-severity action.
NobelActive Internal NP 3.5x15mm Endosseus dental implant, REF: 34128, Rx Only, Sterile R… was recalled by Nobel Biocare Usa Llc in June 10, 2020. Reason: The top label for dental implant may incorrectly indicate the implant length. Specifically, the label may con…. Check the official notice for the remedy. Verify recall #Z-2207-2020 with the FDA Devices before acting.
The recall
Nobel Biocare Usa Llc issued this moderate-severity FDA Devices recall-The top label for dental implant may incorrectly indicate the implant length. Specifically, the label may con….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2207-2020 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2207-2020) was formally reported on June 10, 2020, with the manufacturer initiating the action on July 19, 2019. It is classified under Moderate severity (Class II), with a current status of Terminated. Nobel Biocare Usa Llc is listed as the recalling firm, operating out of Yorba Linda, CA. Federal records list the affected scope as 221 implants.
The documented reason for this recall is: The top label for dental implant may incorrectly indicate the implant length. Specifically, the label may contain 11.5mm for the length, which is shorter than the actual length of the implant (15mm). An incorrect sized… Distribution data in the federal record shows the product reached: US: AL, AR, AZ, CA, CO, CT, FL, HI, ID, IL, IN, KY, MA, MS, MI, MO, NC, NH, NV, NY, OK, PA, SC, SD, TX, UT, WA, WI OUS: None. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
221 implants
Related Recalls
6
3 from same agency
NobelActive Internal NP 3.5x15mm Endosseus dental implant, REF: 34128, Rx Only, Sterile R, CE, UDI: (01)07332747010619
The top label for dental implant may incorrectly indicate the implant length. Specifically, the label may contain 11.5mm for the length, which is shorter than the actual length of the implant (15mm). An incorrect sized implant may cause patient discomfort, permanent nerve damage and may require an additional surgery.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2207-2020 |
| Date reported | June 10, 2020 |
| Date initiated | July 19, 2019 |
| Recalling firm | Nobel Biocare Usa Llc |
| Firm location | Yorba Linda, CA |
| Affected scope | 221 implants |
| Distribution | US: AL, AR, AZ, CA, CO, CT, FL, HI, ID, IL, IN, KY, MA, MS, MI, MO, NC, NH, NV, NY, OK, PA, SC, SD, TX, UT, WA, WI OUS: None |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported June 10, 2020. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.