PlainRecalls
FDA Devices Moderate Class II Terminated

NobelActive Internal NP 3.5x15mm Endosseus dental implant, REF: 34128, Rx Only, Sterile R, CE, UDI: (01)07332747010619

Reported: June 10, 2020 Initiated: July 19, 2019 #Z-2207-2020

Product Description

NobelActive Internal NP 3.5x15mm Endosseus dental implant, REF: 34128, Rx Only, Sterile R, CE, UDI: (01)07332747010619

Reason for Recall

The top label for dental implant may incorrectly indicate the implant length. Specifically, the label may contain 11.5mm for the length, which is shorter than the actual length of the implant (15mm). An incorrect sized implant may cause patient discomfort, permanent nerve damage and may require an additional surgery.

Details

Recalling Firm
Nobel Biocare Usa Llc
Units Affected
221 implants
Distribution
US: AL, AR, AZ, CA, CO, CT, FL, HI, ID, IL, IN, KY, MA, MS, MI, MO, NC, NH, NV, NY, OK, PA, SC, SD, TX, UT, WA, WI OUS: None
Location
Yorba Linda, CA

Frequently Asked Questions

What product was recalled?
NobelActive Internal NP 3.5x15mm Endosseus dental implant, REF: 34128, Rx Only, Sterile R, CE, UDI: (01)07332747010619. Recalled by Nobel Biocare Usa Llc. Units affected: 221 implants.
Why was this product recalled?
The top label for dental implant may incorrectly indicate the implant length. Specifically, the label may contain 11.5mm for the length, which is shorter than the actual length of the implant (15mm). An incorrect sized implant may cause patient discomfort, permanent nerve damage and may require an additional surgery.
Which agency issued this recall?
This recall was issued by the FDA Devices on June 10, 2020. Severity: Moderate. Recall number: Z-2207-2020.