PlainRecalls
ModerateClass IIOngoing

FDA Devices recall · Reported May 27, 2026

OneLIF Interbody Cage (REF/Description): 010-107-1007/Small 7 Degree 10mm X 7mm, 010-107-1007-2/Small 7 Degree 10mm X 7mm - Oblique, 010-107-1209/Small 7 Degree 12mm X 9mm, 010-107-1411/Small 7 Degree 14mm X 11mm, 010-107-1613/Small 7 Degree 16mm X 13mm, 010-113-1006/Small, 13 Degree 10mm X 6mm, 010-113-1006-2/Small, 13 Degree, 10mm X 6mm - Oblique, 010-113-1208/Small 13 Degree 12mm X 8mm, 010-113-1410/Small 13 Degree 14mm X 10mm, 010-113-1612/Small 13 Degree 16mm X 12mm, 010-119-1407

Intervertebral body fusion system straight inserter failed to properly attach to affected cages, which may result in an inability to securely engage the implant with the inserter,…

Recall #
Z-2213-2026
Affected scope
1259
Initiated
April 23, 2026
Verify with FDA Devices →
View my saved recalls

The recall

Novapproach Spine, LLC issued this moderate-severity FDA Devices recall — Intervertebral body fusion system straight inserter failed to properly attach to affected cages, which may re….

Moderate
severity level
1K units
affected scope
Class II
classification
May 27, 2026
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-2213-2026 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This FDA Devices action (record #Z-2213-2026) was formally reported on May 27, 2026, with the manufacturer initiating the action on April 23, 2026. It is classified under Moderate severity (Class II), with a current status of Ongoing. Novapproach Spine, LLC is listed as the recalling firm, operating out of Alachua, FL. Federal records list the affected scope as 1259.

The documented reason for this recall is: Intervertebral body fusion system straight inserter failed to properly attach to affected cages, which may result in an inability to securely engage the implant with the inserter, which could lead to procedural delay an… Distribution data in the federal record shows the product reached: US Nationwide distribution in the states of NY, CT, CA, TX, FL, MI.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices — clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category — 40,409 medical devices recalls on record

-1,00001,0002,0003,0004,0005,000 20052008201120142017202020232026 1,243

Where this recall sits in the database

Severity2366872097High severity (most serious)Moderate severityLow severity
Where this recall sits in the database

Of 100,165 recalls in the database, 23,668 are high severity, 72,097 moderate, and 4,400 low. This recall is classified moderate severity.

Severity

Moderate

Affected scope

1259

Related Recalls

6

6 from same agency

Product description

OneLIF Interbody Cage (REF/Description): 010-107-1007/Small 7 Degree 10mm X 7mm, 010-107-1007-2/Small 7 Degree 10mm X 7mm - Oblique, 010-107-1209/Small 7 Degree 12mm X 9mm, 010-107-1411/Small 7 Degree 14mm X 11mm, 010-107-1613/Small 7 Degree 16mm X 13mm, 010-113-1006/Small, 13 Degree 10mm X 6mm, 010-113-1006-2/Small, 13 Degree, 10mm X 6mm - Oblique, 010-113-1208/Small 13 Degree 12mm X 8mm, 010-113-1410/Small 13 Degree 14mm X 10mm, 010-113-1612/Small 13 Degree 16mm X 12mm, 010-119-1407/19 Degree 14mm X 7mm, 010-119-1609/Small 19 Degree 16mm X 9mm, 010-119-1811/Small 19 Degree 18mm X 11mm, 010-125-1606/Small 25 Degree 16mm X 6mm, 010-125-1808/Small 25 Degree 18mm X 8mm, 010-207-1007/Medium 7 Degree 10mm X 7mm, 010-207-1007-2/Medium 7 Degree 10mm X 7mm - Oblique, 010-207-1209/Medium 7 Degree 12mm X 9mm, 010-207-1411/Medium 7 Degree 14mm X 11mm, 010-207-1613/Medium 7 Degree 16mm X 13mm, 010-213-1206/Medium 13 Degree 12mm X 6mm, 010-213-1408/Medium 13 Degree 14mm X 8mm, 010-213-1610/Medium 13 Degree 16mm X 10mm, 010-213-1812/Medium 13 Degree 18mm X 12mm, 010-219-1406/Medium 19 Degree 14mm X 6mm, 010-219-1608/Medium 19 Degree 16mm X 8mm, 010-219-1810/Medium 19 Degree 18mm X 10mm, 010-225-1807/Medium 25 Degree 18mm X 7mm, 010-225-2009/Medium 25 Degree 20mm X 9mm, 010-307-1006/Large 7 Degree,10mm X 6mm, 010-307-1006-2/Large 7 Degree 10mm X 6mm - Oblique, 010-307-1208/ Large 7 Degree 12mm X 8mm, 010-307-1410/Large 7 Degree 14mm X 10mm, 010-307-1612/Large 7 Degree 16mm X 12mm, 010-313-1205/Large 13 Degree 12mm X 5mm, 010-313-1407/Large 13 Degree 14mm X 7mm, 010-313-1609/Large 13 Degree 16mm X 9mm, 010-313-1811/Large 13 Degree 18mm X 11mm, 010-313-2013/Large 13 Degree 20mm X 13mm, 010-319-1606/Large 19 Degree 16mm X 6mm, 010-319-1808/ Large 19 Degree 18mm X 8mm, 010-319-2010/ Large 19 Degree 20mm X 10mm, 010-325-2007/Large 25 Degree 20mm X 7mm, 010-325-2209/ Large 25 Degree 22mm X 9mm, Implant Kits containing OneLIF Interbody Cages, Name/Kit Family: Kit-OneLIF-10 mm (716)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (717)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (718)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (719)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (722)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (723)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (724)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (733)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (734)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (737)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (738)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (746)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (751)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (761)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (762)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (787)/Kit-OneLIF-10 mm, Kit-OneLIF- IMPOneLIF Implant Kit (701)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (702)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (703) NL/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (704)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (705)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (706)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (708)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (709)/Kit-OneLIF-IMP - OneLIF Implant Kit,

Reason for recall

Intervertebral body fusion system straight inserter failed to properly attach to affected cages, which may result in an inability to securely engage the implant with the inserter, which could lead to procedural delay and/or the need to select an alternative implant.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Ongoing
Recall number Z-2213-2026
Date reported May 27, 2026
Date initiated April 23, 2026
Recalling firm Novapproach Spine, LLC
Firm location Alachua, FL
Affected scope 1259
Distribution US Nationwide distribution in the states of NY, CT, CA, TX, FL, MI.

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

1259 units affected — limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units)
Large-scale (100K – 1M units)
Massive (≥1M units)

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-2213-2026) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting — recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
OneLIF Interbody Cage (REF/Description): 010-107-1007/Small 7 Degree 10mm X 7mm, 010-107-1007-2/Small 7 Degree 10mm X 7mm - Oblique, 010-107-1209/Small 7 Degree 12mm X 9mm, 010-107-1411/Small 7 Degree 14mm X 11mm, 010-107-1613/Small 7 Degree 16mm X 13mm, 010-113-1006/Small, 13 Degree 10mm X 6mm, 010-113-1006-2/Small, 13 Degree, 10mm X 6mm - Oblique, 010-113-1208/Small 13 Degree 12mm X 8mm, 010-113-1410/Small 13 Degree 14mm X 10mm, 010-113-1612/Small 13 Degree 16mm X 12mm, 010-119-1407/19 Degree 14mm X 7mm, 010-119-1609/Small 19 Degree 16mm X 9mm, 010-119-1811/Small 19 Degree 18mm X 11mm, 010-125-1606/Small 25 Degree 16mm X 6mm, 010-125-1808/Small 25 Degree 18mm X 8mm, 010-207-1007/Medium 7 Degree 10mm X 7mm, 010-207-1007-2/Medium 7 Degree 10mm X 7mm - Oblique, 010-207-1209/Medium 7 Degree 12mm X 9mm, 010-207-1411/Medium 7 Degree 14mm X 11mm, 010-207-1613/Medium 7 Degree 16mm X 13mm, 010-213-1206/Medium 13 Degree 12mm X 6mm, 010-213-1408/Medium 13 Degree 14mm X 8mm, 010-213-1610/Medium 13 Degree 16mm X 10mm, 010-213-1812/Medium 13 Degree 18mm X 12mm, 010-219-1406/Medium 19 Degree 14mm X 6mm, 010-219-1608/Medium 19 Degree 16mm X 8mm, 010-219-1810/Medium 19 Degree 18mm X 10mm, 010-225-1807/Medium 25 Degree 18mm X 7mm, 010-225-2009/Medium 25 Degree 20mm X 9mm, 010-307-1006/Large 7 Degree,10mm X 6mm, 010-307-1006-2/Large 7 Degree 10mm X 6mm - Oblique, 010-307-1208/ Large 7 Degree 12mm X 8mm, 010-307-1410/Large 7 Degree 14mm X 10mm, 010-307-1612/Large 7 Degree 16mm X 12mm, 010-313-1205/Large 13 Degree 12mm X 5mm, 010-313-1407/Large 13 Degree 14mm X 7mm, 010-313-1609/Large 13 Degree 16mm X 9mm, 010-313-1811/Large 13 Degree 18mm X 11mm, 010-313-2013/Large 13 Degree 20mm X 13mm, 010-319-1606/Large 19 Degree 16mm X 6mm, 010-319-1808/ Large 19 Degree 18mm X 8mm, 010-319-2010/ Large 19 Degree 20mm X 10mm, 010-325-2007/Large 25 Degree 20mm X 7mm, 010-325-2209/ Large 25 Degree 22mm X 9mm, Implant Kits containing OneLIF Interbody Cages, Name/Kit Family: Kit-OneLIF-10 mm (716)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (717)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (718)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (719)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (722)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (723)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (724)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (733)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (734)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (737)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (738)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (746)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (751)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (761)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (762)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (787)/Kit-OneLIF-10 mm, Kit-OneLIF- IMPOneLIF Implant Kit (701)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (702)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (703) NL/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (704)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (705)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (706)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (708)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (709)/Kit-OneLIF-IMP - OneLIF Implant Kit,. Recalled by Novapproach Spine, LLC. Units affected: 1259.
Why was this product recalled?
Intervertebral body fusion system straight inserter failed to properly attach to affected cages, which may result in an inability to securely engage the implant with the inserter, which could lead to procedural delay and/or the need to select an alternative implant.
Which agency issued this recall?
This recall was issued by the FDA Devices on May 27, 2026. Severity: Moderate. Recall number: Z-2213-2026.
Where was the recalled product distributed?
Distribution: US Nationwide distribution in the states of NY, CT, CA, TX, FL, MI..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-2213-2026) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L… →

Data Sources

Source: FDA, CPSC, and NHTSA federal recall databases. This recall: FDA Devices, reported May 27, 2026.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records — no number is typed in by an editor. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of June 2026.