Severity
Moderate
FDA Devices recall · Reported August 29, 2012
In the recalled lot, the products were not machined according to the production router, a size 52 cup (122136052) was machined instead. This resulted in a size 52 liner being etc…
DePuy Orthopaedics, Inc. recalled Altrx" Altralinked" Polyethylene liners Description: An acetabular liner is one of the… - a moderate-severity action.
Altrx" Altralinked" Polyethylene liners Description: An acetabular liner is one of the… was recalled by DePuy Orthopaedics, Inc. in August 29, 2012. Reason: In the recalled lot, the products were not machined according to the production router, a size 52 cup (122136…. Check the official notice for the remedy. Verify recall #Z-2220-2012 with the FDA Devices before acting.
The recall
DePuy Orthopaedics, Inc. issued this moderate-severity FDA Devices recall-In the recalled lot, the products were not machined according to the production router, a size 52 cup (122136….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2220-2012 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2220-2012) was formally reported on August 29, 2012, with the manufacturer initiating the action on July 24, 2012. It is classified under Moderate severity (Class II), with a current status of Terminated. DePuy Orthopaedics, Inc. is listed as the recalling firm, operating out of Warsaw, IN. Federal records list the affected scope as 20 liners.
The documented reason for this recall is: In the recalled lot, the products were not machined according to the production router, a size 52 cup (122136052) was machined instead. This resulted in a size 52 liner being etched and labeled as a size 54 liner. Distribution data in the federal record shows the product reached: Worldwide Distribution - USA including PA and the country of Canada. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
20 liners
Related Recalls
6
3 from same agency
Altrx" Altralinked" Polyethylene liners Description: An acetabular liner is one of the implants used during total hip arthroplasty to replace the damaged hip joint to help increase patient mobility and to reduce pain. Packaging: Product is placed into a PETG tray, sealed with a Tyvek lid. This is placed into a second PETG tray and sealed with a Tyvek lid. This package is placed into a paperboard carton, labeled, and shrink-wrapped. Intended Use: An acetabular liner is used in combination with other implants for total hip arthroplasty is intended to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components.
In the recalled lot, the products were not machined according to the production router, a size 52 cup (122136052) was machined instead. This resulted in a size 52 liner being etched and labeled as a size 54 liner.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2220-2012 |
| Date reported | August 29, 2012 |
| Date initiated | July 24, 2012 |
| Recalling firm | DePuy Orthopaedics, Inc. |
| Firm location | Warsaw, IN |
| Affected scope | 20 liners |
| Distribution | Worldwide Distribution - USA including PA and the country of Canada |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported August 29, 2012. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.