Severity
Moderate
FDA Devices recall · Reported June 14, 2017
a gap between the conveyor belt and the rack tray table (plastic part) may cause vibration of the 5 position racks during the transport. This issue can pose a potential risk to i…
Roche Diagnostics Corporation recalled Cobas connection module (CCM) - OUTPUT UNIT The module CCM enables a unidirectional conn… - a moderate-severity action.
Cobas connection module (CCM) - OUTPUT UNIT The module CCM enables a unidirectional conn… was recalled by Roche Diagnostics Corporation in June 14, 2017. Reason: a gap between the conveyor belt and the rack tray table (plastic part) may cause vibration of the 5 position …. Check the official notice for the remedy. Verify recall #Z-2223-2017 with the FDA Devices before acting.
The recall
Roche Diagnostics Corporation issued this moderate-severity FDA Devices recall-a gap between the conveyor belt and the rack tray table (plastic part) may cause vibration of the 5 position ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2223-2017 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2223-2017) was formally reported on June 14, 2017, with the manufacturer initiating the action on May 9, 2017. It is classified under Moderate severity (Class II), with a current status of Terminated. Roche Diagnostics Corporation is listed as the recalling firm, operating out of Indianapolis, IN. Federal records list the affected scope as 3.
The documented reason for this recall is: a gap between the conveyor belt and the rack tray table (plastic part) may cause vibration of the 5 position racks during the transport. This issue can pose a potential risk to involved operators and can also lead to p… Distribution data in the federal record shows the product reached: US distribution to WA only.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
3
Related Recalls
6
3 from same agency
Cobas connection module (CCM) - OUTPUT UNIT The module CCM enables a unidirectional connection of the pre-analytical systems cobas p 612/ p 512 to analysis instruments from Roche Diagnostics, Hitachi or Sysmex by means of the Aloka transportation system
a gap between the conveyor belt and the rack tray table (plastic part) may cause vibration of the 5 position racks during the transport. This issue can pose a potential risk to involved operators and can also lead to potential cross contamination of samples in the affected racks.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2223-2017 |
| Date reported | June 14, 2017 |
| Date initiated | May 9, 2017 |
| Recalling firm | Roche Diagnostics Corporation |
| Firm location | Indianapolis, IN |
| Affected scope | 3 |
| Distribution | US distribution to WA only. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported June 14, 2017. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.