Cannulated graduated drill bit 1.9 mm diameter, with AO QC system, REF: INS-114 for use in orthopedic surgery.

Reported: August 18, 2021 Initiated: November 3, 2020 #Z-2236-2021

Product Description

Cannulated graduated drill bit 1.9 mm diameter, with AO QC system, REF: INS-114 for use in orthopedic surgery.

Reason for Recall

Traces of corrosion on the laser marking was detected on cannulated graduated drill bit, which could lead to the emission of wear / corrosion particles in the operated area, potentially leading to the following risks: Inflammation, pain, allergy, adverse tissue reaction. Device description: The drill bit is used to create the hole required for the introduction of the screw, it is suitable for Neosteo compression screws for cortical and spongy bone. Drilling depth can be read on the instrument. The shaft is coupled with the ¿4.5mm surgical motor using a AO ¿4.5mm male drive connexion. The countersink bit is coupled to a ¿1.0mm k-wire to ensure guiding while drilling. The instrument is packaged in a plastic sheath sealed by welding.

Details

Recalling Firm
NEOSTEO
Units Affected
25
Distribution
US Nationwide distribution in the state of New Jersey.
Location
Reze, N/A

Frequently Asked Questions

What product was recalled?
Cannulated graduated drill bit 1.9 mm diameter, with AO QC system, REF: INS-114 for use in orthopedic surgery.. Recalled by NEOSTEO. Units affected: 25.
Why was this product recalled?
Traces of corrosion on the laser marking was detected on cannulated graduated drill bit, which could lead to the emission of wear / corrosion particles in the operated area, potentially leading to the following risks: Inflammation, pain, allergy, adverse tissue reaction. Device description: The drill bit is used to create the hole required for the introduction of the screw, it is suitable for Neosteo compression screws for cortical and spongy bone. Drilling depth can be read on the instrument. The shaft is coupled with the ¿4.5mm surgical motor using a AO ¿4.5mm male drive connexion. The countersink bit is coupled to a ¿1.0mm k-wire to ensure guiding while drilling. The instrument is packaged in a plastic sheath sealed by welding.
Which agency issued this recall?
This recall was issued by the FDA Devices on August 18, 2021. Severity: Moderate. Recall number: Z-2236-2021.