BD Trucount Controls, 30 Test - REF: 340335

Reported: August 13, 2025 Initiated: July 8, 2025 #Z-2238-2025

Product Description

BD Trucount Controls, 30 Test - REF: 340335

Reason for Recall

Control kits contain incorrect standard deviation (SD) values for the low and high control beads which compromises the accurate calculation of bead counts.

Details

Units Affected
629 kits
Distribution
Worldwide - US Nationwide distribution in the states of Tennessee, New York, Massachusetts and the countries of Argentina, Belgium, Botswana, Brazil, Canada, France, Germany, Lithuania, Peru, Philippines, United Kingdom.
Location
Milpitas, CA

Frequently Asked Questions

What product was recalled?
BD Trucount Controls, 30 Test - REF: 340335. Recalled by Becton, Dickinson and Company, BD Bio Sciences. Units affected: 629 kits.
Why was this product recalled?
Control kits contain incorrect standard deviation (SD) values for the low and high control beads which compromises the accurate calculation of bead counts.