FDA Devices recall · Reported June 3, 2026
Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in c…
Abiomed, Inc. recalled Automated Impella Controller (AIC) with the below product descriptions and corresponding … - a critical-severity action.
Automated Impella Controller (AIC) with the below product descriptions and corresponding … was recalled by Abiomed, Inc. in June 3, 2026. Reason: Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-…. Check the official notice for the remedy. Verify recall #Z-2238-2026 with the FDA Devices before acting.
The recall
Abiomed, Inc. issued this critical-severity FDA Devices recall-Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2238-2026 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2238-2026) was formally reported on June 3, 2026, with the manufacturer initiating the action on January 17, 2011. It is classified under Critical severity (Class I), with a current status of Ongoing. Abiomed, Inc. is listed as the recalling firm, operating out of Danvers, MA. Federal records list the affected scope as 625 units.
The documented reason for this recall is: Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is impro… Distribution data in the federal record shows the product reached: Worldwide distribution - US Nationwide and the countries of Canada, Denmark, France, Germany, Italy, Kuwait, Netherlands, Norway, Saudi Arabia, Spain, Switzerland, United Arab Emirates, United Kingdom.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified high severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Critical
Affected scope
625 units
Related Recalls
6
3 from same agency
Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US; Product Code: 0042-0000-US.
Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper alignment between the purge cassette and the motor drive in the AIC. This can result in a piston block event and lead to the inability to complete priming of the purge system, requiring the user to switch to backup AIC console.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Critical (Class I) |
| Status | Ongoing |
| Recall number | Z-2238-2026 |
| Date reported | June 3, 2026 |
| Date initiated | January 17, 2011 |
| Recalling firm | Abiomed, Inc. |
| Firm location | Danvers, MA |
| Affected scope | 625 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Canada, Denmark, France, Germany, Italy, Kuwait, Netherlands, Norway, Saudi Arabia, Spain, Switzerland, United Arab Emirates, United Kingdom. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported June 3, 2026. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.