Fabius GS Premium Anesthesia Machine; Cat. no. 8607000 Product Usage: Inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series machines are equipped with a compact breathing system; providing fresh gas decoupling, PEEP, and pressure limitation. The following ventilation options are available: Volume Controlled Ventilation Pressure Co

Reported: June 27, 2018 Initiated: May 24, 2018 #Z-2250-2018

Product Description

Fabius GS Premium Anesthesia Machine; Cat. no. 8607000 Product Usage: Inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series machines are equipped with a compact breathing system; providing fresh gas decoupling, PEEP, and pressure limitation. The following ventilation options are available: Volume Controlled Ventilation Pressure Controlled Ventilation Pressure Support (Optional) SIMV/PS (Optional) Manual Ventilation Spontaneous Breathing Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2). As per IEC 60601-2-13 (Anesthetic Workstations and their Modules-Particular Requirements), additional monitoring of the concentrations of CO2 and anesthetic agent is required when the machine is in use. IEC 60601-2-13:2003 requires that a manual ventilation bag must be available for emergency use. Fresh gas enrichment is provided by the Dr¿ger Vapor anesthetic vaporizer. Fabius MRI can also be used in MRI scanner rooms with magnets 1.5 tesla and 3 tesla at a fringe field strength of 40 mtesla or less

Reason for Recall

The Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would not prevent setting an N2O flow that would result in a hypoxic mixture from being dosed to the patient. Potential adverse outcomes include death of the patient.

Details

Recalling Firm
Draeger Medical, Inc.
Units Affected
62
Distribution
US Nationwide Distribution - Healthcare facilities in AK, AL, AZ, CA, CO, FL, GA, IA, IL, IN, MA, MD, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, WA, WI, and WV.
Location
Telford, PA

Frequently Asked Questions

What product was recalled?
Fabius GS Premium Anesthesia Machine; Cat. no. 8607000 Product Usage: Inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series machines are equipped with a compact breathing system; providing fresh gas decoupling, PEEP, and pressure limitation. The following ventilation options are available: Volume Controlled Ventilation Pressure Controlled Ventilation Pressure Support (Optional) SIMV/PS (Optional) Manual Ventilation Spontaneous Breathing Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2). As per IEC 60601-2-13 (Anesthetic Workstations and their Modules-Particular Requirements), additional monitoring of the concentrations of CO2 and anesthetic agent is required when the machine is in use. IEC 60601-2-13:2003 requires that a manual ventilation bag must be available for emergency use. Fresh gas enrichment is provided by the Dr¿ger Vapor anesthetic vaporizer. Fabius MRI can also be used in MRI scanner rooms with magnets 1.5 tesla and 3 tesla at a fringe field strength of 40 mtesla or less. Recalled by Draeger Medical, Inc.. Units affected: 62.
Why was this product recalled?
The Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would not prevent setting an N2O flow that would result in a hypoxic mixture from being dosed to the patient. Potential adverse outcomes include death of the patient.
Which agency issued this recall?
This recall was issued by the FDA Devices on June 27, 2018. Severity: Moderate. Recall number: Z-2250-2018.