Severity
Moderate
FDA Devices recall · Reported June 27, 2018
The Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would not prevent setting an N2O flow that would result in a hypoxic mixture…
Draeger Medical, Inc. recalled Apollo Anesthesia Machine; Cat. no. 8605310 Product Usage: Indicated as a continuous … - a moderate-severity action.
Apollo Anesthesia Machine; Cat. no. 8605310 Product Usage: Indicated as a continuous … was recalled by Draeger Medical, Inc. in June 27, 2018. Reason: The Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would n…. Check the official notice for the remedy. Verify recall #Z-2253-2018 with the FDA Devices before acting.
The recall
Draeger Medical, Inc. issued this moderate-severity FDA Devices recall-The Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would n….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2253-2018 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2253-2018) was formally reported on June 27, 2018, with the manufacturer initiating the action on May 24, 2018. It is classified under Moderate severity (Class II), with a current status of Terminated. Draeger Medical, Inc. is listed as the recalling firm, operating out of Telford, PA. Federal records list the affected scope as 72.
The documented reason for this recall is: The Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would not prevent setting an N2O flow that would result in a hypoxic mixture from being dosed to the patient. Potent… Distribution data in the federal record shows the product reached: US Nationwide Distribution - Healthcare facilities in AK, AL, AZ, CA, CO, FL, GA, IA, IL, IN, MA, MD, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, WA, WI, and WV.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
72
Related Recalls
6
3 from same agency
Apollo Anesthesia Machine; Cat. no. 8605310 Product Usage: Indicated as a continuous flow anesthesia system. The Apollo may be used for manually assisted or automatic ventilation, delivery of gases and anesthetic vapor, and monitoring of oxygen and CO2 concentration, breathing pressure, respiratory volume, and anesthetic agent concentration and identification. It is an inhalation anesthesia machine for use in operating, induction, and recovery rooms. It can be used with rebreathing systems, semiclosed to virtually closed systems with low flow and minimal flow techniques, and non-rebreathing systems (with the Auxiliary Common Gas Outlet). It may be used with O2, N2O, and Air supplied by a medical gas pipeline system or by externally mounted gas cylinders. Anesthetic agent can be delivered via vaporizers mounted on the machine. The Apollo is equipped with a compact breathing system, providing fresh-gas decoupling, PEEP, and pressure limitation. It has an electrically driven and electronically controlled ventilator.
The Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would not prevent setting an N2O flow that would result in a hypoxic mixture from being dosed to the patient. Potential adverse outcomes include death of the patient.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2253-2018 |
| Date reported | June 27, 2018 |
| Date initiated | May 24, 2018 |
| Recalling firm | Draeger Medical, Inc. |
| Firm location | Telford, PA |
| Affected scope | 72 |
| Distribution | US Nationwide Distribution - Healthcare facilities in AK, AL, AZ, CA, CO, FL, GA, IA, IL, IN, MA, MD, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, WA, WI, and WV. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported June 27, 2018. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.