FDA Devices recall · Reported October 2, 2013
Terumo Pall AL6X Arterial Blood Line Filter, labeled in part Terumo Cardiovascular Systems Corporation, packaged as part of a Terumo Cardiovascular Procedure Kit or as a single, sterile filter unit Single/Sterile Terumo Pall AL6X Arterial Filter: The Terumo Pall AL6X Arterial Filter is indicated for use in cardiopulmonary bypass procedures for the removal of micro-emboli greater than 40 microns in size, including gas emboli, fat emboli and aggregates composed of platelets, red blood cells and
Medical device used in cardiovascular procedures is defective.
Terumo Cardiovascular Systems Corporation recalled Terumo Pall AL6X Arterial Blood Line Filter, labeled in part Terumo Cardiovascular System… - a moderate-severity action.
Terumo Pall AL6X Arterial Blood Line Filter, labeled in part Terumo Cardiovascular System… was recalled by Terumo Cardiovascular Systems Corporation in October 2, 2013. Reason: Medical device used in cardiovascular procedures is defective.. Check the official notice for the remedy. Verify recall #Z-2261-2013 with the FDA Devices before acting.
The recall
Terumo Cardiovascular Systems Corporation issued this moderate-severity FDA Devices recall-Medical device used in cardiovascular procedures is defective..
- Moderate
- severity level
- Class II
- classification
- October 2, 2013
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-2261-2013 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
Terumo Pall AL6X Arterial Blood Line Filter, labeled in part Terumo Cardiovascular Systems Corporation, packaged as part of a Terumo Cardiovascular Procedure Kit or as a single, sterile filter unit Single/Sterile Terumo Pall AL6X Arterial Filter: The Terumo Pall AL6X Arterial Filter is indicated for use in cardiopulmonary bypass procedures for the removal of micro-emboli greater than 40 microns in size, including gas emboli, fat emboli and aggregates composed of platelets, red blood cells and other debris from the arterial line and where the flow rate will not exceed 8 liters per minute. Cardiovascular Procedure (Convenience) Kit that includes the Terumo Pall AL6X Arterial Filter: The Cardiovascular Procedure Kits are indicated for use only in the extracorporeal circuit for the cardiopulmonary bypass procedures for which the user designed. The sterile cardiovascular procedure kit is intended to be used one time for periods up to 6 hours.
Reason for recall
Medical device used in cardiovascular procedures is defective.
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2261-2013 |
| Date reported | October 2, 2013 |
| Date initiated | July 26, 2013 |
| Recalling firm | Terumo Cardiovascular Systems Corporation |
| Firm location | Elkton, MD |
| Affected scope | 15,817 filters |
| Distribution | Nationwide Distribution including WA, IN, OK, TX, MI, CA, NY, CT, AZ, FL, WV, NC, VA, PA, TN, GA, AL, MA, SC, OH, MS, MO, SD, WI, AR, AND NV. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-2261-2013) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls
Methodology · Corrections · Archive as of August 2026