PlainRecalls
FDA Devices Verify with FDA Devices → Moderate Class II Terminated

Cisco brand HealthPresence, version 2.1; Model/Catalog Number: CHP-ENTPRSYR2.I-K9; Product is manufactured and distributed by Cisco Systems, Inc. San Jose, CA Cisco HealthPresence, version 2 is intended for remote consultation of patients with a healthcare provider. Cisco HealthPresence enables remote collaboration between a physician and a patient by providing: high-definition video and advanced audio; transmission and display of third-party medical device data; integration with third party

Reported: September 5, 2012 Initiated: July 23, 2012 #Z-2263-2012 1 unit units

Cisco Systems Inc issued this FDA Devices recall on September 5, 2012. Classified as Moderate severity (Class II). Approximately 1 unit units are affected. The recall was issued because: Software anomaly may result in incorrect values and interpretations. The initial vital signs reading will display accur…. This recall notice is sourced from official FDA Devices enforcement records. Below you will find the complete product description, hazard information, remedy instructions, and related recalls from the same manufacturer or product category.

Recall Insight

This FDA Devices action (record #Z-2263-2012) was formally reported on September 5, 2012, with the manufacturer initiating the action on July 23, 2012. It is classified under Moderate severity (Class II), with a current status of Terminated. Cisco Systems Inc is listed as the recalling firm, operating out of San Jose, CA. Federal records indicate 1 unit units are affected.

The documented reason for this recall is: Software anomaly may result in incorrect values and interpretations. The initial vital signs reading will display accurately on the Cisco Health Presence Vitals tab and the OnePlace tab. However, subsequent vital signs … Distribution data in the federal record shows the product reached: Nationwide Distribution including the state of TN.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

To put this record in context, PlainRecalls indexes 83,949 recalls across the FDA, CPSC, NHTSA and USDA FSIS going back to 1995. Within the same product category, the database holds 6 closely related recalls, of which 6 were also issued by FDA Devices. That clustering is a signal — repeated actions in a narrow category often indicate a systemic quality-control issue, a supplier-wide contamination, or a design defect that has propagated across product lines. This recall is roughly 14 years old; older recalls can remain relevant because many units enter resale, rental, and secondary-market channels where the original warning never reaches the end user. Always cross-check the recall number against the official agency page before relying on any summary.

Recall Distribution by Severity Class

Severity1Class I (Critical)Class II (Moderate)Class III (Low)
Recall Distribution by Severity Class

Severity

Moderate

Units Affected

1 unit

Related Recalls

6

6 from same agency

Product Description

Cisco brand HealthPresence, version 2.1; Model/Catalog Number: CHP-ENTPRSYR2.I-K9; Product is manufactured and distributed by Cisco Systems, Inc. San Jose, CA Cisco HealthPresence, version 2 is intended for remote consultation of patients with a healthcare provider. Cisco HealthPresence enables remote collaboration between a physician and a patient by providing: high-definition video and advanced audio; transmission and display of third-party medical device data; integration with third party EMR: scheduling: and other software applications for remote consultations and exams. Cisco Health Presence is not intended for use in emergency situations. Cisco HealthPresence is not intended for use in active patient monitoring or alarming.

Reason for Recall

Software anomaly may result in incorrect values and interpretations. The initial vital signs reading will display accurately on the Cisco Health Presence Vitals tab and the OnePlace tab. However, subsequent vital signs readings will only update on the Cisco HealthPresence Vitals tab and within the OnePlace database, but will not update under the OnePlace tab, and thus, will not provide current vit

Details

Recalling Firm
Cisco Systems Inc
Units Affected
1 unit
Distribution
Nationwide Distribution including the state of TN.
Location
San Jose, CA

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-2263-2012
Date reported September 5, 2012
Date initiated July 23, 2012
Recalling firm Cisco Systems Inc
Units affected 1 unit
Distribution Nationwide Distribution including the state of TN.

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

1 unit units affected — limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units)
Large-scale (100K – 1M units)
Massive (≥1M units)

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

Frequently Asked Questions

What product was recalled?
Cisco brand HealthPresence, version 2.1; Model/Catalog Number: CHP-ENTPRSYR2.I-K9; Product is manufactured and distributed by Cisco Systems, Inc. San Jose, CA Cisco HealthPresence, version 2 is intended for remote consultation of patients with a healthcare provider. Cisco HealthPresence enables remote collaboration between a physician and a patient by providing: high-definition video and advanced audio; transmission and display of third-party medical device data; integration with third party EMR: scheduling: and other software applications for remote consultations and exams. Cisco Health Presence is not intended for use in emergency situations. Cisco HealthPresence is not intended for use in active patient monitoring or alarming.. Recalled by Cisco Systems Inc. Units affected: 1 unit.
Why was this product recalled?
Software anomaly may result in incorrect values and interpretations. The initial vital signs reading will display accurately on the Cisco Health Presence Vitals tab and the OnePlace tab. However, subsequent vital signs readings will only update on the Cisco HealthPresence Vitals tab and within the OnePlace database, but will not update under the OnePlace tab, and thus, will not provide current vit
Which agency issued this recall?
This recall was issued by the FDA Devices on September 5, 2012. Severity: Moderate. Recall number: Z-2263-2012.
Where was the recalled product distributed?
Distribution: Nationwide Distribution including the state of TN..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-2263-2012) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Recall Context

Product recalls are issued when a manufacturer, distributor, or federal agency determines that a product poses a safety risk to consumers. This recall is classified as moderate severity, indicating the product may cause temporary or medically reversible health consequences. Across PlainRecalls, we track 83,000+ recalls from FDA, CPSC, and NHTSA to help consumers stay informed and act quickly when safety issues arise.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Endo-Model Replacement Plateau; Item Number: 15-0027/11; →

Data Sources

Data as of 2025. Source: FDA, CPSC, NHTSA, USDA FSIS federal recall databases.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)
  • Source: USDA FSIS — Food Safety and Inspection Service (meat, poultry, and egg product recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

All federal data sources used on this page

Source: Federal recall agencies (FDA, CPSC, NHTSA, USDA FSIS) Aggregated multi-agency recall feeds · 2024 Recall data normalized across federal agency feeds; severity classifications follow each agency's own taxonomy (FDA Class I/II/III; CPSC, NHTSA, USDA FSIS).