Ovation iX Abdominal Stent Graft Systems, REF: TV-AB2080-I, TV-AB2080-J, TV-AB2380-I, TV-AB2380-J, TV-AB2680-I, TV-AB2680-J, TV-AB2980-I, TV-AB2980-J, TV-AB3480-I, TV-AB3480-J
Reported: June 17, 2020 Initiated: May 6, 2020 #Z-2263-2020
Product Description
Ovation iX Abdominal Stent Graft Systems, REF: TV-AB2080-I, TV-AB2080-J, TV-AB2380-I, TV-AB2380-J, TV-AB2680-I, TV-AB2680-J, TV-AB2980-I, TV-AB2980-J, TV-AB3480-I, TV-AB3480-J
Reason for Recall
A material weakness adjacent to the polymer fill channel may become compromised during pressurization with liquid polymer. Polymer leaks may occur during implantation of abdominal stent graft system. Clinical events related to polymer leaks may be systemic and/or aneurysm related.
Details
- Recalling Firm
- Endologix, Inc.
- Units Affected
- 9156
- Distribution
- Worldwide DIstribution: U.S (nationwide).: PA, AZ, NY, FL, HI, CA, OR, IL, MS, AR, WI, KY, TX, TN, MO, NJ, MA, MT, VA, MI, CO, WV, NC, NH, NV, CT, ME, WY, OH, OK, IN, UT, AL, WA, GA, SC, KS, LA, NM, MD, DC, NE, SD, RI, IA, VT, ND. and Countries of:: Argentina, Australia, Chile, Hong Kong, Japan, Malaysia, Mexico, New Zealand, Taiwan, Thailand, Canada, Austria, Belgium, Cyprus, Czech Republic, Germany, Hungary, Ireland, Italy, Switzerland, Luxembourg, Netherlands, Poland, Portugal, Romania, Spain, Sweden, and United Kingdom
- Location
- Santa Rosa, CA
Frequently Asked Questions
What product was recalled? ▼
Ovation iX Abdominal Stent Graft Systems, REF: TV-AB2080-I, TV-AB2080-J, TV-AB2380-I, TV-AB2380-J, TV-AB2680-I, TV-AB2680-J, TV-AB2980-I, TV-AB2980-J, TV-AB3480-I, TV-AB3480-J. Recalled by Endologix, Inc.. Units affected: 9156.
Why was this product recalled? ▼
A material weakness adjacent to the polymer fill channel may become compromised during pressurization with liquid polymer. Polymer leaks may occur during implantation of abdominal stent graft system. Clinical events related to polymer leaks may be systemic and/or aneurysm related.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on June 17, 2020. Severity: Critical. Recall number: Z-2263-2020.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11