FDA Devices recall · Reported September 5, 2012
Whole Body X--ray Scanner This device is indicated as a whole body volume CT system. The system produces volumes of up to 320 axial slices in a single rotation.
1. It has been found that in raw data processing for Go&Return helical scanning, when the raw data acquired in the first scan (Go) and that acquired in the second scan (Return) ar…
Toshiba American Medical Systems Inc recalled Whole Body X--ray Scanner This device is indicated as a whole body volume CT system. T… - a moderate-severity action.
Whole Body X--ray Scanner This device is indicated as a whole body volume CT system. T… was recalled by Toshiba American Medical Systems Inc in September 5, 2012. Reason: 1. It has been found that in raw data processing for Go&Return helical scanning, when the raw data acquired i…. Check the official notice for the remedy. Verify recall #Z-2282-2012 with the FDA Devices before acting.
The recall
Toshiba American Medical Systems Inc issued this moderate-severity FDA Devices recall-1. It has been found that in raw data processing for Go&Return helical scanning, when the raw data acquired i….
- Moderate
- severity level
- Class II
- classification
- September 5, 2012
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-2282-2012 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
Whole Body X--ray Scanner This device is indicated as a whole body volume CT system. The system produces volumes of up to 320 axial slices in a single rotation.
Reason for recall
1. It has been found that in raw data processing for Go&Return helical scanning, when the raw data acquired in the first scan (Go) and that acquired in the second scan (Return) are selected and reconstructed at one time, incorrect couch positions and scan times are displayed on the reconstructed images for the data acquired by Return scanning, while all related information is displayed on the reco
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2282-2012 |
| Date reported | September 5, 2012 |
| Date initiated | July 13, 2009 |
| Recalling firm | Toshiba American Medical Systems Inc |
| Firm location | Tustin, CA |
| Affected scope | 24 Systems |
| Distribution | Nationwide distribution: USA including states of: AL, AR, CA, CO, FL, GA, KY, MA, MD, MT, ND, NJ, NV, NY, OH, PA and Puerto Rico. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-2282-2012) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls
Methodology · Corrections · Archive as of August 2026