Atellica CH Enzymatic Creatinine_2, Siemens Material Number 11097533, UDI Number 00630414595795 - Product Usage: use in the quantitative determination of creatinine in human serum, plasma, and urine for the diagnosis and treatment of renal disease.
Reported: June 17, 2020 Initiated: May 4, 2020 #Z-2283-2020
Product Description
Atellica CH Enzymatic Creatinine_2, Siemens Material Number 11097533, UDI Number 00630414595795 - Product Usage: use in the quantitative determination of creatinine in human serum, plasma, and urine for the diagnosis and treatment of renal disease.
Reason for Recall
The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.
Details
- Recalling Firm
- Siemens Healthcare Diagnostics, Inc.
- Units Affected
- 1488
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of North Carolina, California, Indiana, Florida, Ohio, Arizona, Massachusetts, New Jersey, New Hampshire, Illinois, Minnesota, Arkansas, New York, Delaware, Nevada, Kentucky, Washington DC, Montana, Missouri, Colorado, Maryland, Tennessee, Oregon, Virginia, Iowa, Wisconsin, Louisiana, Georgia, Pennsylvania, North Dakota, Mississippi, South Dakota, Washington, Nebraska, Texas, Alaska, Puerto Rico, Idaho, Oregon, Maine, Vermont, West Virginia, Kansas.
- Location
- Newark, DE
Frequently Asked Questions
What product was recalled? ▼
Atellica CH Enzymatic Creatinine_2, Siemens Material Number 11097533, UDI Number 00630414595795 - Product Usage: use in the quantitative determination of creatinine in human serum, plasma, and urine for the diagnosis and treatment of renal disease.. Recalled by Siemens Healthcare Diagnostics, Inc.. Units affected: 1488.
Why was this product recalled? ▼
The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on June 17, 2020. Severity: Moderate. Recall number: Z-2283-2020.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11