Willscher Vas Reapproximator Plate, Surgical Instrument, Supplied sterilized by ethylene oxide gas in peel-open packages. Used to stabilize the vas ends during microscopic vasovasostomy.
Reported: June 14, 2017 Initiated: May 1, 2017 #Z-2312-2017
Product Description
Willscher Vas Reapproximator Plate, Surgical Instrument, Supplied sterilized by ethylene oxide gas in peel-open packages. Used to stabilize the vas ends during microscopic vasovasostomy.
Reason for Recall
COOK Medical is initiating a voluntary recall of multiple products because the reprocessing instructions do not provide sufficient detailed information for the cleaning, disinfection, and sterilization of these products.
Details
- Recalling Firm
- Cook Inc.
- Units Affected
- 327 lots
- Distribution
- Worldwide Distribution - US (nationwide) and countries of: Australia, Austria, Belgium, Bahrain, Canada, China, Costa Rica, Finland, France, Germany, Hong Kong, India, Ireland, Italy, Lithuania, Latvia, Malaysia, Netherlands, Romania, Russia, Saudi Arabia, Singapore, Spain, Switzerland, United Arab Emirates, and United Kingdom.
- Location
- Bloomington, IN
Frequently Asked Questions
What product was recalled? ▼
Willscher Vas Reapproximator Plate, Surgical Instrument, Supplied sterilized by ethylene oxide gas in peel-open packages. Used to stabilize the vas ends during microscopic vasovasostomy.. Recalled by Cook Inc.. Units affected: 327 lots.
Why was this product recalled? ▼
COOK Medical is initiating a voluntary recall of multiple products because the reprocessing instructions do not provide sufficient detailed information for the cleaning, disinfection, and sterilization of these products.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on June 14, 2017. Severity: Moderate. Recall number: Z-2312-2017.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11