Description/REF: BD PYXIS MEDBANK MINI 1FH-1FM/169-137, BD PYXIS MEDBANK MINI 4HM/169-113, BD PYXIS MEDBANK MINI 4HH/169-112, BD PYXIS MEDBANK MINI 3HH-1HM/169-111, BD PYXIS MEDBANK MINI 3HH-1FG/169-110, BD PYXIS MEDBANK MINI 2HM-1FM/169-109, BD PYXIS MEDBANK MINI 2HH-2HM/169-108, BD PYXIS MEDBANK MINI 2HH-1FM/169-107, BD PYXIS MEDBANK MINI 2HH-1FH/169-106, BD PYXIS MEDBANK MINI 2FM/169-105, BD PYXIS MEDBANK MINI 2FH/169-104, BD PYXIS MEDBANK MINI 1HH-3HM/169-103, BD PYXIS

Reported: August 20, 2025 Initiated: July 10, 2025 #Z-2321-2025 6076 units

CareFusion 303, Inc. issued this FDA Devices recall on August 20, 2025. Classified as Low severity (Class III). Approximately 6076 units are affected. The recall was issued because: Reports of delayed access to medication in automated dispensing cabinets because of the override/Add Item workflow.. This recall notice is sourced from official FDA Devices enforcement records. Below you will find the complete product description, hazard information, remedy instructions, and related recalls from the same manufacturer or product category.

Recall Insight

This FDA Devices action (record #Z-2321-2025) was formally reported on August 20, 2025, with the manufacturer initiating the action on July 10, 2025. It is classified under Low severity (Class III), with a current status of Ongoing. CareFusion 303, Inc. is listed as the recalling firm, operating out of San Diego, CA. Federal records indicate 6076 units are affected.

The documented reason for this recall is: Reports of delayed access to medication in automated dispensing cabinets because of the override/Add Item workflow. Distribution data in the federal record shows the product reached: US Nationwide distribution including in the states of NY, MA, KY, VA, OR, OH, FL, WA, MS, OK, TN, PA, RI, TX, SC, AR, MI, NJ, CT, MD, AZ, CA, WV, NC, UT, WY, CO, SD, ID, IL, IN, NV, NM, NE, KS, GA, WI, MO, IA, ME, LA, M…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

To put this record in context, PlainRecalls indexes 83,949 recalls across the FDA, CPSC, NHTSA and USDA FSIS going back to 1995. Within the same product category, the database holds 6 closely related recalls, of which 6 were also issued by FDA Devices. That clustering is a signal — repeated actions in a narrow category often indicate a systemic quality-control issue, a supplier-wide contamination, or a design defect that has propagated across product lines. This recall is roughly 1 year old; older recalls can remain relevant because many units enter resale, rental, and secondary-market channels where the original warning never reaches the end user. Always cross-check the recall number against the official agency page before relying on any summary.

Recall Distribution by Severity Class

Severity1Class I (Critical)Class II (Moderate)Class III (Low)
Recall Distribution by Severity Class

Severity

Low

Units Affected

6076

Related Recalls

6

6 from same agency

Recall Progress (industry avg ~60%) 60.0%

Product Description

Description/REF: BD PYXIS MEDBANK MINI 1FH-1FM/169-137, BD PYXIS MEDBANK MINI 4HM/169-113, BD PYXIS MEDBANK MINI 4HH/169-112, BD PYXIS MEDBANK MINI 3HH-1HM/169-111, BD PYXIS MEDBANK MINI 3HH-1FG/169-110, BD PYXIS MEDBANK MINI 2HM-1FM/169-109, BD PYXIS MEDBANK MINI 2HH-2HM/169-108, BD PYXIS MEDBANK MINI 2HH-1FM/169-107, BD PYXIS MEDBANK MINI 2HH-1FH/169-106, BD PYXIS MEDBANK MINI 2FM/169-105, BD PYXIS MEDBANK MINI 2FH/169-104, BD PYXIS MEDBANK MINI 1HH-3HM/169-103, BD PYXIS MEDBANK MINI 1HH-1HM-1FM/169-102, BD PYXIS MEDBANK MINI 1HH-1FH-1HM/169-101 , BD PYXIS MEDBANK MINI 1FH-2HH/155311-01, BD PYXIS MEDBANK MINI 4HH/155310-01, BD PYXIS MEDBANK MINI 1FH-1FM/155288-01, BD PYXIS MEDBANK MINI 4HM/155171-01, BD PYXIS MEDBANK MINI 3HH-1HM/155169-01, BD PYXIS MEDBANK MINI 3HH-1FG/155168-01, BD PYXIS MEDBANK MINI 2HM-1FM/155167-01, BD PYXIS MEDBANK MINI 2HH-2HM/155166-01, BD PYXIS MEDBANK MINI 2HH-1FM/155165-01, BD PYXIS MEDBANK MINI 2FM/155163-01, BD PYXIS MEDBANK MINI 1HH-3HM/155161-01, BD PYXIS MEDBANK MINI 1HH-1HM-1FM/155160-01, BD PYXIS MEDBANK MINI 1HH-1FH-1HM/155159-01, BD PYXIS MEDBANK MINI 2HM-1FM/139131-01, BD PYXIS MEDBANK MINI 1HH-1HM-1FM/139122-01, BD PYXIS MEDBANK MINI 3HH-1FG/139091-01, BD PYXIS MEDBANK MINI 1HH-1FH-1HM/138921-01, BD PYXIS MEDBANK MINI 4HM/138915-01, BD PYXIS MEDBANK MINI 1HH-3HM/138912-01, BD PYXIS MEDBANK MINI 4HH/138910-01, BD PYXIS MEDBANK MINI 3HH-1HM/138909-01, BD PYXIS MEDBANK MINI 2HH-1FH/138908-01, BD PYXIS MEDBANK MINI 2HH-2HM/138907-01, BD PYXIS MEDBANK MINI 2HH-1FM/138906-01, BD PYXIS MEDBANK MINI 2FM/138903-01, BD PYXIS MEDBANK MINI 2FH/138902-01, MEDBANK MINI MAIN/1147-00, MEDBANK MAIN/1137-00, BD PYXIS MEDBANK TWR MN 8HH-3FM/169-81, BD PYXIS MEDBANK TWR MN 8HH-3FH/169-80, BD PYXIS MEDBANK TWR MN 8HH-3FF/169-79, BD PYXIS MEDBANK TWR MN 8HH-2HM-2FM/169-78, BD PYXIS MEDBANK TWR MN 8HH-2FH-1FM/169-77, BD PYXIS MEDBANK TWR MN 8HH-1FH-2HM-1FM/169-76, BD PYXIS MEDBANK TWR MN 8HH-1FH-2FM/169-75, BD PYXIS MEDBANK TWR MN 7HH-1HM-3FM/169-74, BD PYXIS MEDBANK TWR MN 7FM/169-73, BD PYXIS MEDBANK TWR MN 7FH/169-72, BD PYXIS MEDBANK TWR MN 6HH-8HM/169-71, BD PYXIS MEDBANK TWR MN 6HH-6HM-1FM/169-70, BD PYXIS MEDBANK TWR MN 6HH-4HM-2FM/169-69, BD PYXIS MEDBANK TWR MN 6HH-4FM/169-68, BD PYXIS MEDBANK TWR MN 6HH-4FH /169-67, BD PYXIS MEDBANK TWR MN 6HH-2FH-2HM-1FM/169-66, BD PYXIS MEDBANK TWR MN 6HH-2FH-2FM/169-65, BD PYXIS MEDBANK TWR MN 6HH-1FH-3FM/169-64, BD PYXIS MEDBANK TWR MN 6HH-1FH-2HM-2FM/169-63, BD PYXIS MEDBANK TWR MN 6FH-1FM/169-62, BD PYXIS MEDBANK TWR MN 5HH-9HM/169-61, BD PYXIS MEDBANK TWR MN 5HH-3HM-3FM/169-60, BD PYXIS MEDBANK TWR MN 4HM-5FM/169-59, BD PYXIS MEDBANK TWR MN 4HH-8HM-1FM/169-58, BD PYXIS MEDBANK TWR MN 4HH-6HM-2FM/169-57, BD PYXIS MEDBANK TWR MN 4HH-5FM/169-56, BD PYXIS MEDBANK TWR MN 4HH-5FH /169-55, BD PYXIS MEDBANK TWR MN 4HH-4HM-3FM/169-54, BD PYXIS MEDBANK TWR MN 4HH-4FH-1FM/169-53, BD PYXIS MEDBANK TWR MN 4HH-3FH-2FM/169-52, BD PYXIS MEDBANK TWR MN 4HH-2HM-4FM/169-51, BD PYXIS MEDBANK TWR MN 4HH-2FH-3FM/169-50, BD PYXIS MEDBANK TWR MN 4HH-2FH-2HM-2FM/169-49, BD PYXIS MEDBANK TWR MN 4HH-1FH-6HM-1FM/169-48, BD PYXIS MEDBANK TWR MN 4HH-1FH-4HM-2FM/169-47, BD PYXIS MEDBANK TWR MN 4HH-1FH-4FM/169-46, BD PYXIS MEDBANK TWR MN 4HH-1FH-2HM-3FM/169-45, BD PYXIS MEDBANK TWR MN 4HH-10HM/169-44, BD PYXIS MEDBANK TWR MN 3HH-7HM-2FM/169-43, BD PYXIS MEDBANK TWR MN 3FH-4FM/169-42, BD PYXIS MEDBANK TWR MN 2HM-6FM/169-41, BD PYXIS MEDBANK TWR MN 2HH-8HM-2FM/169-40, BD PYXIS MEDBANK TWR MN 2HH-6HM-3FM/169-39, BD PYXIS MEDBANK TWR MN 2HH-6FM/169-38, BD PYXIS MEDBANK TWR MN 2HH-4HM-4FM/169-37, BD PYXIS MEDBANK TWR MN 2HH-4HM-4FF/169-36, BD PYXIS MEDBANK TWR MN 2HH-2FM-4FF/169-35, BD PYXIS MEDBANK TWR MN 2HH-2FH-4FM/169-34, BD PYXIS MEDBANK TWR MN 2HH-1FH-5FM/169-33, BD PYXIS MEDBANK TWR MN 2HH-10HM-1FM/169-32, BD PY

Reason for Recall

Reports of delayed access to medication in automated dispensing cabinets because of the override/Add Item workflow.

Details

Recalling Firm
CareFusion 303, Inc.
Units Affected
6076
Distribution
US Nationwide distribution including in the states of NY, MA, KY, VA, OR, OH, FL, WA, MS, OK, TN, PA, RI, TX, SC, AR, MI, NJ, CT, MD, AZ, CA, WV, NC, UT, WY, CO, SD, ID, IL, IN, NV, NM, NE, KS, GA, WI, MO, IA, ME, LA, MN, AL, VT, NH, DE, MT, ND, AK, HI, DC.
Location
San Diego, CA

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Low (Class III)
Status Ongoing
Recall number Z-2321-2025
Date reported August 20, 2025
Date initiated July 10, 2025
Recalling firm CareFusion 303, Inc.
Units affected 6076
Distribution US Nationwide distribution including in the states of NY, MA, KY, VA, OR, OH, FL, WA, MS, OK, TN, PA, RI, TX, SC, AR, MI, NJ, CT, MD, AZ, CA, WV, NC, UT, WY, CO, SD, ID, IL, IN, NV, NM, NE, KS, GA, WI, MO, IA, ME, LA, MN, AL, VT, NH, DE, M…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

6076 units affected — limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units)
Large-scale (100K – 1M units)
Massive (≥1M units)

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

Frequently Asked Questions

What product was recalled?
Description/REF: BD PYXIS MEDBANK MINI 1FH-1FM/169-137, BD PYXIS MEDBANK MINI 4HM/169-113, BD PYXIS MEDBANK MINI 4HH/169-112, BD PYXIS MEDBANK MINI 3HH-1HM/169-111, BD PYXIS MEDBANK MINI 3HH-1FG/169-110, BD PYXIS MEDBANK MINI 2HM-1FM/169-109, BD PYXIS MEDBANK MINI 2HH-2HM/169-108, BD PYXIS MEDBANK MINI 2HH-1FM/169-107, BD PYXIS MEDBANK MINI 2HH-1FH/169-106, BD PYXIS MEDBANK MINI 2FM/169-105, BD PYXIS MEDBANK MINI 2FH/169-104, BD PYXIS MEDBANK MINI 1HH-3HM/169-103, BD PYXIS MEDBANK MINI 1HH-1HM-1FM/169-102, BD PYXIS MEDBANK MINI 1HH-1FH-1HM/169-101 , BD PYXIS MEDBANK MINI 1FH-2HH/155311-01, BD PYXIS MEDBANK MINI 4HH/155310-01, BD PYXIS MEDBANK MINI 1FH-1FM/155288-01, BD PYXIS MEDBANK MINI 4HM/155171-01, BD PYXIS MEDBANK MINI 3HH-1HM/155169-01, BD PYXIS MEDBANK MINI 3HH-1FG/155168-01, BD PYXIS MEDBANK MINI 2HM-1FM/155167-01, BD PYXIS MEDBANK MINI 2HH-2HM/155166-01, BD PYXIS MEDBANK MINI 2HH-1FM/155165-01, BD PYXIS MEDBANK MINI 2FM/155163-01, BD PYXIS MEDBANK MINI 1HH-3HM/155161-01, BD PYXIS MEDBANK MINI 1HH-1HM-1FM/155160-01, BD PYXIS MEDBANK MINI 1HH-1FH-1HM/155159-01, BD PYXIS MEDBANK MINI 2HM-1FM/139131-01, BD PYXIS MEDBANK MINI 1HH-1HM-1FM/139122-01, BD PYXIS MEDBANK MINI 3HH-1FG/139091-01, BD PYXIS MEDBANK MINI 1HH-1FH-1HM/138921-01, BD PYXIS MEDBANK MINI 4HM/138915-01, BD PYXIS MEDBANK MINI 1HH-3HM/138912-01, BD PYXIS MEDBANK MINI 4HH/138910-01, BD PYXIS MEDBANK MINI 3HH-1HM/138909-01, BD PYXIS MEDBANK MINI 2HH-1FH/138908-01, BD PYXIS MEDBANK MINI 2HH-2HM/138907-01, BD PYXIS MEDBANK MINI 2HH-1FM/138906-01, BD PYXIS MEDBANK MINI 2FM/138903-01, BD PYXIS MEDBANK MINI 2FH/138902-01, MEDBANK MINI MAIN/1147-00, MEDBANK MAIN/1137-00, BD PYXIS MEDBANK TWR MN 8HH-3FM/169-81, BD PYXIS MEDBANK TWR MN 8HH-3FH/169-80, BD PYXIS MEDBANK TWR MN 8HH-3FF/169-79, BD PYXIS MEDBANK TWR MN 8HH-2HM-2FM/169-78, BD PYXIS MEDBANK TWR MN 8HH-2FH-1FM/169-77, BD PYXIS MEDBANK TWR MN 8HH-1FH-2HM-1FM/169-76, BD PYXIS MEDBANK TWR MN 8HH-1FH-2FM/169-75, BD PYXIS MEDBANK TWR MN 7HH-1HM-3FM/169-74, BD PYXIS MEDBANK TWR MN 7FM/169-73, BD PYXIS MEDBANK TWR MN 7FH/169-72, BD PYXIS MEDBANK TWR MN 6HH-8HM/169-71, BD PYXIS MEDBANK TWR MN 6HH-6HM-1FM/169-70, BD PYXIS MEDBANK TWR MN 6HH-4HM-2FM/169-69, BD PYXIS MEDBANK TWR MN 6HH-4FM/169-68, BD PYXIS MEDBANK TWR MN 6HH-4FH /169-67, BD PYXIS MEDBANK TWR MN 6HH-2FH-2HM-1FM/169-66, BD PYXIS MEDBANK TWR MN 6HH-2FH-2FM/169-65, BD PYXIS MEDBANK TWR MN 6HH-1FH-3FM/169-64, BD PYXIS MEDBANK TWR MN 6HH-1FH-2HM-2FM/169-63, BD PYXIS MEDBANK TWR MN 6FH-1FM/169-62, BD PYXIS MEDBANK TWR MN 5HH-9HM/169-61, BD PYXIS MEDBANK TWR MN 5HH-3HM-3FM/169-60, BD PYXIS MEDBANK TWR MN 4HM-5FM/169-59, BD PYXIS MEDBANK TWR MN 4HH-8HM-1FM/169-58, BD PYXIS MEDBANK TWR MN 4HH-6HM-2FM/169-57, BD PYXIS MEDBANK TWR MN 4HH-5FM/169-56, BD PYXIS MEDBANK TWR MN 4HH-5FH /169-55, BD PYXIS MEDBANK TWR MN 4HH-4HM-3FM/169-54, BD PYXIS MEDBANK TWR MN 4HH-4FH-1FM/169-53, BD PYXIS MEDBANK TWR MN 4HH-3FH-2FM/169-52, BD PYXIS MEDBANK TWR MN 4HH-2HM-4FM/169-51, BD PYXIS MEDBANK TWR MN 4HH-2FH-3FM/169-50, BD PYXIS MEDBANK TWR MN 4HH-2FH-2HM-2FM/169-49, BD PYXIS MEDBANK TWR MN 4HH-1FH-6HM-1FM/169-48, BD PYXIS MEDBANK TWR MN 4HH-1FH-4HM-2FM/169-47, BD PYXIS MEDBANK TWR MN 4HH-1FH-4FM/169-46, BD PYXIS MEDBANK TWR MN 4HH-1FH-2HM-3FM/169-45, BD PYXIS MEDBANK TWR MN 4HH-10HM/169-44, BD PYXIS MEDBANK TWR MN 3HH-7HM-2FM/169-43, BD PYXIS MEDBANK TWR MN 3FH-4FM/169-42, BD PYXIS MEDBANK TWR MN 2HM-6FM/169-41, BD PYXIS MEDBANK TWR MN 2HH-8HM-2FM/169-40, BD PYXIS MEDBANK TWR MN 2HH-6HM-3FM/169-39, BD PYXIS MEDBANK TWR MN 2HH-6FM/169-38, BD PYXIS MEDBANK TWR MN 2HH-4HM-4FM/169-37, BD PYXIS MEDBANK TWR MN 2HH-4HM-4FF/169-36, BD PYXIS MEDBANK TWR MN 2HH-2FM-4FF/169-35, BD PYXIS MEDBANK TWR MN 2HH-2FH-4FM/169-34, BD PYXIS MEDBANK TWR MN 2HH-1FH-5FM/169-33, BD PYXIS MEDBANK TWR MN 2HH-10HM-1FM/169-32, BD PY. Recalled by CareFusion 303, Inc.. Units affected: 6076.
Why was this product recalled?
Reports of delayed access to medication in automated dispensing cabinets because of the override/Add Item workflow.
Which agency issued this recall?
This recall was issued by the FDA Devices on August 20, 2025. Severity: Low. Recall number: Z-2321-2025.
Where was the recalled product distributed?
Distribution: US Nationwide distribution including in the states of NY, MA, KY, VA, OR, OH, FL, WA, MS, OK, TN, PA, RI, TX, SC, AR, MI, NJ, CT, MD, AZ, CA, WV, NC, UT, WY, CO, SD, ID, IL, IN, NV, NM, NE, KS, GA, WI, MO, IA, ME, LA, MN, AL, VT, NH, DE, MT, ND, AK, HI, DC..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-2321-2025) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Recall Context

Product recalls are issued when a manufacturer, distributor, or federal agency determines that a product poses a safety risk to consumers. Recall severity varies based on the likelihood and degree of potential harm to consumers. Across PlainRecalls, we track 83,000+ recalls from FDA, CPSC, and NHTSA to help consumers stay informed and act quickly when safety issues arise.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Endo-Model Replacement Plateau; Item Number: 15-0027/11; →

Data Sources

Data as of 2025. Source: FDA, CPSC, NHTSA, USDA FSIS federal recall databases.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)
  • Source: USDA FSIS — Food Safety and Inspection Service (meat, poultry, and egg product recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

All federal data sources used on this page

Source: Federal recall agencies (FDA, CPSC, NHTSA, USDA FSIS) Aggregated multi-agency recall feeds · 2024 Recall data normalized across federal agency feeds; severity classifications follow each agency's own taxonomy (FDA Class I/II/III; CPSC, NHTSA, USDA FSIS).