Severity
Critical
FDA Devices recall · Reported August 27, 2025
Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replace…
Boston Scientific Corporation recalled ENDOTAK RELIANCE G/SG with 4-SITE Connector, transvenous defibrillation lead, Permanent D… - a critical-severity action.
ENDOTAK RELIANCE G/SG with 4-SITE Connector, transvenous defibrillation lead, Permanent D… was recalled by Boston Scientific Corporation in August 27, 2025. Reason: Affected lead models may experience gradually rising LVSI measurements in association with calcification of t…. Check the official notice for the remedy. Verify recall #Z-2326-2025 with the FDA Devices before acting.
The recall
Boston Scientific Corporation issued this critical-severity FDA Devices recall-Affected lead models may experience gradually rising LVSI measurements in association with calcification of t….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2326-2025 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2326-2025) was formally reported on August 27, 2025, with the manufacturer initiating the action on July 24, 2025. It is classified under Critical severity (Class I), with a current status of Ongoing. Boston Scientific Corporation is listed as the recalling firm, operating out of Saint Paul, MN. Federal records list the affected scope as 247,702 US; 344,670 OUS.
The documented reason for this recall is: Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need… Distribution data in the federal record shows the product reached: Worldwide Distribution. US nationwide. International distribution worldwide.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified high severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Critical
Affected scope
247,702 US; 344,670 OUS
Related Recalls
6
3 from same agency
ENDOTAK RELIANCE G/SG with 4-SITE Connector, transvenous defibrillation lead, Permanent Defibrillator Electrodes; Model Nos. 0282, 0283, 0285, 0286, 0292, 0293, 0295, 0296
Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Critical (Class I) |
| Status | Ongoing |
| Recall number | Z-2326-2025 |
| Date reported | August 27, 2025 |
| Date initiated | July 24, 2025 |
| Recalling firm | Boston Scientific Corporation |
| Firm location | Saint Paul, MN |
| Affected scope | 247,702 US; 344,670 OUS |
| Distribution | Worldwide Distribution. US nationwide. International distribution worldwide. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported August 27, 2025. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.