Severity
Moderate
FDA Devices recall · Reported July 17, 2024
There were 27 complaints between March 22, 2024, and April 30, 2024, related to VH-3500 and VH-4000, reporting that the C-Ring wire was straightened (due to insufficient bend radi…
Maquet Cardiovascular, LLC recalled The VASOVIEW HEMOPRO Endoscopic Vessel Harvesting Systems is designed for use in conjunct… - a moderate-severity action.
The VASOVIEW HEMOPRO Endoscopic Vessel Harvesting Systems is designed for use in conjunct… was recalled by Maquet Cardiovascular, LLC in July 17, 2024. Reason: There were 27 complaints between March 22, 2024, and April 30, 2024, related to VH-3500 and VH-4000, reportin…. Check the official notice for the remedy. Verify recall #Z-2337-2024 with the FDA Devices before acting.
The recall
Maquet Cardiovascular, LLC issued this moderate-severity FDA Devices recall-There were 27 complaints between March 22, 2024, and April 30, 2024, related to VH-3500 and VH-4000, reportin….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2337-2024 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2337-2024) was formally reported on July 17, 2024, with the manufacturer initiating the action on May 31, 2024. It is classified under Moderate severity (Class II), with a current status of Ongoing. Maquet Cardiovascular, LLC is listed as the recalling firm, operating out of Wayne, NJ. Federal records list the affected scope as 600 units.
The documented reason for this recall is: There were 27 complaints between March 22, 2024, and April 30, 2024, related to VH-3500 and VH-4000, reporting that the C-Ring wire was straightened (due to insufficient bend radius) and that the distance between C-Ring… Distribution data in the federal record shows the product reached: NationwideForeign: AT DE DK ES GB HK IT NL ZA. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
600 units
Related Recalls
6
3 from same agency
The VASOVIEW HEMOPRO Endoscopic Vessel Harvesting Systems is designed for use in conjunction with the 7 mm Extended Length Endoscope. The Harvesting Cannula has four lumens which house the Endoscope, C-Ring, distal lens washer tube, and harvesting tool for cutting and cauterizing vessel branches. The harvesting tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning of the device, cutting, and cauterization are performed under endoscopic visualization.
There were 27 complaints between March 22, 2024, and April 30, 2024, related to VH-3500 and VH-4000, reporting that the C-Ring wire was straightened (due to insufficient bend radius) and that the distance between C-Ring and Harvesting Tool was closer than normal. No adverse events have been reported to date in direct association with this issue.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-2337-2024 |
| Date reported | July 17, 2024 |
| Date initiated | May 31, 2024 |
| Recalling firm | Maquet Cardiovascular, LLC |
| Firm location | Wayne, NJ |
| Affected scope | 600 units |
| Distribution | NationwideForeign: AT DE DK ES GB HK IT NL ZA |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported July 17, 2024. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.