PlainRecalls
FDA Devices Moderate Class II Ongoing

GE HealthCare OEC 9800.

Reported: September 3, 2025 Initiated: July 2, 2025 #Z-2341-2025

Product Description

GE HealthCare OEC 9800.

Reason for Recall

GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will lead to a small increase in radiation exposure to the patient and/or operator. The missing lead tape is not detectable by the user.

Details

Recalling Firm
GE Medical Systems, LLC
Distribution
US Nationwide distribution.
Location
Waukesha, WI

Frequently Asked Questions

What product was recalled?
GE HealthCare OEC 9800.. Recalled by GE Medical Systems, LLC.
Why was this product recalled?
GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will lead to a small increase in radiation exposure to the patient and/or operator. The missing lead tape is not detectable by the user.
Which agency issued this recall?
This recall was issued by the FDA Devices on September 3, 2025. Severity: Moderate. Recall number: Z-2341-2025.