FDA Devices Moderate Class II Terminated

91496 Ultraview SL Command Module, Options A, B, C, and I. Option A: ECG/Respiration, SpO2, Temperature x2 & adult/neonate NIBP. Option B: ECG/Respiration, Invasive Pressure x2, SpO2, Temperature x2 & adult/neonate NIBP. Option C: ECG/Respiration, Invasive Pressure x4, SpO2, Cardiac Output, Temperature x2 & adult/neonate NIBP. Option I: SpO2, Temperature x2, & adult/neonate NIBP.

Reported: August 10, 2016 Initiated: July 1, 2016 #Z-2343-2016

Product Description

91496 Ultraview SL Command Module, Options A, B, C, and I. Option A: ECG/Respiration, SpO2, Temperature x2 & adult/neonate NIBP. Option B: ECG/Respiration, Invasive Pressure x2, SpO2, Temperature x2 & adult/neonate NIBP. Option C: ECG/Respiration, Invasive Pressure x4, SpO2, Cardiac Output, Temperature x2 & adult/neonate NIBP. Option I: SpO2, Temperature x2, & adult/neonate NIBP.

Reason for Recall

The firm has received multiple reports of the Non-Invasive Blood Pressure (NIBP) parameter becoming non-functional with an associated message (No Reading). In addition, there is one of the following three error messages: Inflate Error, HW Error, or No Data.

Details

Recalling Firm
Spacelabs Healthcare Inc
Units Affected
3801 modules and 116 PCBAs & valves distributed in the US; 1297 modules and 366 PCBAs & valves distributed outside the US
Distribution
Worldwide Distribution - USA Distribution and to the countries of : AFGHANISTAN, ARGENTINA, AUSTRALIA, BAHRAIN, BOLIVIA, BRAZIL, BRUNEI DARUSSALAM; CANADA, CAYMAN ISLANDS, CHILE, CHINA, COLOMBIA, COSTA RICA, CZECH REPUBLIC, DOMINICAN REPUBLIC; ECUADOR, EGYPT, ESTONIA, FRANCE, GERMANY, GUATEMALA, HONG KONG, INDIA, IRAN (ISLAMIC REPUBLIC OF), ISRAEL, ITALY, JORDAN, KUWAIT, MEXICO, MOROCCO, NETHERLANDS, NICARAGUA, PAKISTAN, PALESTINE, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, SAUDI ARABIA, Slovakia, SPAIN, SRI LANKA, SWITZERLAND, TAIWAN, TRINIDAD AND TOBAGO, TUNISIA, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, and VIET NAM.
Location
Snoqualmie, WA

Frequently Asked Questions

What product was recalled?
91496 Ultraview SL Command Module, Options A, B, C, and I. Option A: ECG/Respiration, SpO2, Temperature x2 & adult/neonate NIBP. Option B: ECG/Respiration, Invasive Pressure x2, SpO2, Temperature x2 & adult/neonate NIBP. Option C: ECG/Respiration, Invasive Pressure x4, SpO2, Cardiac Output, Temperature x2 & adult/neonate NIBP. Option I: SpO2, Temperature x2, & adult/neonate NIBP.. Recalled by Spacelabs Healthcare Inc. Units affected: 3801 modules and 116 PCBAs & valves distributed in the US; 1297 modules and 366 PCBAs & valves distributed outside the US.
Why was this product recalled?
The firm has received multiple reports of the Non-Invasive Blood Pressure (NIBP) parameter becoming non-functional with an associated message (No Reading). In addition, there is one of the following three error messages: Inflate Error, HW Error, or No Data.
Which agency issued this recall?
This recall was issued by the FDA Devices on August 10, 2016. Severity: Moderate. Recall number: Z-2343-2016.