Severity
Critical
Alere San Diego, Inc. issued this FDA Devices recall on August 31, 2016. Classified as Critical severity (Class I). The recall was issued because: Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring Systems (Test Strip…. This recall notice is sourced from official FDA Devices enforcement records. Below you will find the complete product description, hazard information, remedy instructions, and related recalls from the same manufacturer or product category.
This FDA Devices action (record #Z-2360-2016) was formally reported on August 31, 2016, with the manufacturer initiating the action on July 11, 2016. It is classified under Critical severity (Class I), with a current status of Terminated. Alere San Diego, Inc. is listed as the recalling firm, operating out of San Diego, CA. The number of affected units is not disclosed in the agency filing, which is common for drug and food recalls where lot-level tracking supersedes unit counts.
The documented reason for this recall is: Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-… Distribution data in the federal record shows the product reached: Worldwide Distribution - US Nationwide and foreign distribution to the countries of: Argentina, Australia, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Brunei Daruss, Canada, Chile, China, Colombia, Egypt, France, …. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
To put this record in context, PlainRecalls indexes 83,949 recalls across the FDA, CPSC, NHTSA and USDA FSIS going back to 1995. Within the same product category, the database holds 6 closely related recalls, of which 6 were also issued by FDA Devices. That clustering is a signal — repeated actions in a narrow category often indicate a systemic quality-control issue, a supplier-wide contamination, or a design defect that has propagated across product lines. This recall is roughly 10 years old; older recalls can remain relevant because many units enter resale, rental, and secondary-market channels where the original warning never reaches the end user. Always cross-check the recall number against the official agency page before relying on any summary.
Severity
Critical
Units Affected
Unknown
Related Recalls
6
6 from same agency
Alere INRatio2 PT/INR Professional Monitoring System Model Number: 55128A Product Usage: The Alere INRatio¿/INRatio¿2 system uses a modified version of the one-stage PT test. After a drop of blood is applied to the test strip, it is drawn into the test area and mixed with reagents that initiate coagulation. When the blood clots, changes in impedance in the sample are detected by the monitor. The monitor calculates the PT and INR results from this impedance change and reports them on the display. The Alere INRatio/INRatio2 PT/INR Monitoring System is used for the quantitative measurement of Prothrombin Time (PT) in fresh, capillary whole blood. The Alere INRatio/INRatio2 PT/INR Monitoring System is intended for use outside the body (in vitro diagnostic use). The Alere INRatio/INRatio2 PT/INR Monitoring System is intended for professional and home use by people taking warfarin and other oral anticoagulant (blood thinning) therapy who need to monitor the clotting time of their blood. The Alere INRatio/INRatio2 PT/INR Monitoring System is not intended to be used for screening purposes.
Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Critical (Class I) |
| Status | Terminated |
| Recall number | Z-2360-2016 |
| Date reported | August 31, 2016 |
| Date initiated | July 11, 2016 |
| Recalling firm | Alere San Diego, Inc. |
| Units affected | Not disclosed |
| Distribution | Worldwide Distribution - US Nationwide and foreign distribution to the countries of: Argentina, Australia, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Brunei Daruss, Canada, Chile, China, Colombia, Egypt, France, Hong Kong, India, In… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
Learn how the US recall system works and how to protect yourself and your household.
How the US Recall System Works
The three federal agencies, how recalls are initiated, and what happens next.
Understanding Recall Severity Classes
What Class I, II, and III mean and which recalls demand immediate action.
What to Do When a Product Is Recalled
Verify, claim your remedy, report injuries, and navigate the process.
How to Check If Your Products Are Recalled
Step-by-step guide to checking food, products, medications, and vehicles.
Recalled Products in Your Home
A room-by-room household audit guide for active recalls.
Most Recalled Product Categories
Rankings of highest-recall categories from FDA, CPSC, and NHTSA.
Product recalls are issued when a manufacturer, distributor, or federal agency determines that a product poses a safety risk to consumers. This recall is classified as critical severity, meaning the product carries a reasonable probability of serious adverse health consequences or death. Across PlainRecalls, we track 83,000+ recalls from FDA, CPSC, and NHTSA to help consumers stay informed and act quickly when safety issues arise.
Explore related product safety and public health data from federal sources.
Food Safety Inspections
FDA and state food facility inspection results — violations, enforcement actions, and compliance history for food manufacturers and processors.
PlainFoodSafe →
Ingredient Safety Data
FDA food additive and ingredient safety database — regulatory status, usage limits, and safety assessments for thousands of ingredients.
PlainIngredients →
Nutrition Data
USDA nutrition facts for 2M+ food items — calories, macros, vitamins, and ingredient analysis to verify what you consume.
GetFoodFacts →
Drug Safety Information
FDA drug data for 680+ medications — interactions, alternatives, side effects, and safety information for recalled and active drugs.
PlainMeds →
Product Injury Data
CPSC emergency room injury data for 838 product categories — 7.3M NEISS records tracking real-world consumer product injuries.
PlainSafety →
Recall Checker
Search our database of 83K+ recalls by product name, brand, or recall number across all agencies.
Check Now →
Recall Radar
Live feed of the latest recalls across FDA, CPSC, NHTSA, and USDA — filter by agency and severity.
View Radar →
Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.
Waldemar Link GmbH & Co. KG (Mfg Site) · 2026-03-18
Penner Patient Care, Inc. · 2026-03-18
Instrumentation Laboratory · 2026-03-18
Beckman Coulter Mishima K.K. · 2026-03-18
Trividia Health, Inc. · 2026-03-18
Compare this recall with Endo-Model Replacement Plateau; Item Number: 15-0027/11; →
Data as of 2025. Source: FDA, CPSC, NHTSA, USDA FSIS federal recall databases.
Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.
Read our methodology — how this data is sourced, computed, and verified.
Source: Federal recall agencies (FDA, CPSC, NHTSA, USDA FSIS) Aggregated multi-agency recall feeds · 2024 Recall data normalized across federal agency feeds; severity classifications follow each agency's own taxonomy (FDA Class I/II/III; CPSC, NHTSA, USDA FSIS).