Severity
Moderate
FDA Devices recall · Reported July 18, 2018
Intuitive Surgical, Inc. (ISI) has recently become aware of an issue with the da Vinci Xi Stapler 45 Blue and Green Reloads that may be missing a component of the knife mechanism.…
Intuitive Surgical, Inc. recalled da Vinci Xi EndoWrist Stapler 45 Reload Blue, Model 48645B-03, 12 reloads/box. - a moderate-severity action.
da Vinci Xi EndoWrist Stapler 45 Reload Blue, Model 48645B-03, 12 reloads/box. was recalled by Intuitive Surgical, Inc. in July 18, 2018. Reason: Intuitive Surgical, Inc. (ISI) has recently become aware of an issue with the da Vinci Xi Stapler 45 Blue and…. Check the official notice for the remedy. Verify recall #Z-2380-2018 with the FDA Devices before acting.
The recall
Intuitive Surgical, Inc. issued this moderate-severity FDA Devices recall-Intuitive Surgical, Inc. (ISI) has recently become aware of an issue with the da Vinci Xi Stapler 45 Blue and….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2380-2018 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2380-2018) was formally reported on July 18, 2018, with the manufacturer initiating the action on April 23, 2018. It is classified under Moderate severity (Class II), with a current status of Terminated. Intuitive Surgical, Inc. is listed as the recalling firm, operating out of Sunnyvale, CA. Federal records list the affected scope as 7337.
The documented reason for this recall is: Intuitive Surgical, Inc. (ISI) has recently become aware of an issue with the da Vinci Xi Stapler 45 Blue and Green Reloads that may be missing a component of the knife mechanism. Without this component, the reload knif… Distribution data in the federal record shows the product reached: U.S., foreign and government accounts. Foreign: Australia, Austria, Belgium, Brazil, Canada, China (Hong Kong), Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Japan, Luxembourg, Netherlands, Norway,…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,422 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
7337
Related Recalls
6
0 from same agency
da Vinci Xi EndoWrist Stapler 45 Reload Blue, Model 48645B-03, 12 reloads/box.
Intuitive Surgical, Inc. (ISI) has recently become aware of an issue with the da Vinci Xi Stapler 45 Blue and Green Reloads that may be missing a component of the knife mechanism. Without this component, the reload knife will not move and cut tissue when the stapler is fired.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2380-2018 |
| Date reported | July 18, 2018 |
| Date initiated | April 23, 2018 |
| Recalling firm | Intuitive Surgical, Inc. |
| Firm location | Sunnyvale, CA |
| Affected scope | 7337 |
| Distribution | U.S., foreign and government accounts. Foreign: Australia, Austria, Belgium, Brazil, Canada, China (Hong Kong), Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Japan, Luxembourg, Netherlands, Norway, Qatar, Saudi Arabia… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-07-02
· 2026-07-02
· 2026-06-25
· 2026-06-25
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported July 18, 2018. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.