Severity
Moderate
FDA Devices recall · Reported September 26, 2012
Creepage Distance between the two primary phases in the GEO Transformer is 4.5 mm.does not meet (IEC) International standard of 5.5 mm.
Philips Healthcare Inc. recalled Philips DigitalDiagnost Systems with M Cabinet Model: 712025 Stationary radiographic… - a moderate-severity action.
Philips DigitalDiagnost Systems with M Cabinet Model: 712025 Stationary radiographic… was recalled by Philips Healthcare Inc. in September 26, 2012. Reason: Creepage Distance between the two primary phases in the GEO Transformer is 4.5 mm.does not meet (IEC) Intern…. Check the official notice for the remedy. Verify recall #Z-2384-2012 with the FDA Devices before acting.
The recall
Philips Healthcare Inc. issued this moderate-severity FDA Devices recall-Creepage Distance between the two primary phases in the GEO Transformer is 4.5 mm.does not meet (IEC) Intern….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2384-2012 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2384-2012) was formally reported on September 26, 2012, with the manufacturer initiating the action on August 14, 2012. It is classified under Moderate severity (Class II), with a current status of Terminated. Philips Healthcare Inc. is listed as the recalling firm, operating out of Andover, MA. Federal records list the affected scope as 4 systems.
The documented reason for this recall is: Creepage Distance between the two primary phases in the GEO Transformer is 4.5 mm.does not meet (IEC) International standard of 5.5 mm. Distribution data in the federal record shows the product reached: Worldwide Distribution - USA including OH and WA and the country of Canada. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
4 systems
Related Recalls
6
3 from same agency
Philips DigitalDiagnost Systems with M Cabinet Model: 712025 Stationary radiographic system, digital-Universal RAD Applications
Creepage Distance between the two primary phases in the GEO Transformer is 4.5 mm.does not meet (IEC) International standard of 5.5 mm.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2384-2012 |
| Date reported | September 26, 2012 |
| Date initiated | August 14, 2012 |
| Recalling firm | Philips Healthcare Inc. |
| Firm location | Andover, MA |
| Affected scope | 4 systems |
| Distribution | Worldwide Distribution - USA including OH and WA and the country of Canada |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported September 26, 2012. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.