Hitachi Scenaria Computed Tomography System Hitachi Medical Systems America, Twinsburg, OH 44087

Reported: August 17, 2016 Initiated: June 30, 2015 #Z-2392-2016

Product Description

Hitachi Scenaria Computed Tomography System Hitachi Medical Systems America, Twinsburg, OH 44087

Reason for Recall

The firm discovered that the centrifugal force applied to the power supply may cause it to fail over time. The exam performed at the time of the failure will likely be lost, resulting in the need to rescan the patient.

Details

Units Affected
33 units
Distribution
US Distribution to the states of : CA, FL, GA, IA, IL, IN, MD, NC, NY, OH, PA, SD, TN, TX & WY., and Internationally to Brazil and Mexico.
Location
Twinsburg, OH

Frequently Asked Questions

What product was recalled?
Hitachi Scenaria Computed Tomography System Hitachi Medical Systems America, Twinsburg, OH 44087. Recalled by Hitachi Medical Systems America Inc. Units affected: 33 units.
Why was this product recalled?
The firm discovered that the centrifugal force applied to the power supply may cause it to fail over time. The exam performed at the time of the failure will likely be lost, resulting in the need to rescan the patient.
Which agency issued this recall?
This recall was issued by the FDA Devices on August 17, 2016. Severity: Moderate. Recall number: Z-2392-2016.