Hitachi Scenaria Computed Tomography System Hitachi Medical Systems America, Twinsburg, OH 44087
Reported: August 17, 2016 Initiated: June 30, 2015 #Z-2392-2016
Product Description
Hitachi Scenaria Computed Tomography System Hitachi Medical Systems America, Twinsburg, OH 44087
Reason for Recall
The firm discovered that the centrifugal force applied to the power supply may cause it to fail over time. The exam performed at the time of the failure will likely be lost, resulting in the need to rescan the patient.
Details
- Recalling Firm
- Hitachi Medical Systems America Inc
- Units Affected
- 33 units
- Distribution
- US Distribution to the states of : CA, FL, GA, IA, IL, IN, MD, NC, NY, OH, PA, SD, TN, TX & WY., and Internationally to Brazil and Mexico.
- Location
- Twinsburg, OH
Frequently Asked Questions
What product was recalled? ▼
Hitachi Scenaria Computed Tomography System Hitachi Medical Systems America, Twinsburg, OH 44087. Recalled by Hitachi Medical Systems America Inc. Units affected: 33 units.
Why was this product recalled? ▼
The firm discovered that the centrifugal force applied to the power supply may cause it to fail over time. The exam performed at the time of the failure will likely be lost, resulting in the need to rescan the patient.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on August 17, 2016. Severity: Moderate. Recall number: Z-2392-2016.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11