TRUMATCH CT CUT GUIDE KIT L- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation to assist in the positioning of a joint replacement component intra-operatively and in guiding the marking of bone before cutting Part Number: 420915
Reported: September 8, 2021 Initiated: August 10, 2021 #Z-2398-2021
Product Description
TRUMATCH CT CUT GUIDE KIT L- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation to assist in the positioning of a joint replacement component intra-operatively and in guiding the marking of bone before cutting Part Number: 420915
Reason for Recall
Software: A coding error associated with the Fast3D Segmentation software. During the scanning process the three images required of the hip, knee and ankle may not be aligned, causing the anatomic landmark locations to potentially result in limb malalignment.
Details
- Recalling Firm
- DePuy Orthopaedics, Inc.
- Units Affected
- 5 cases ( 1/cs)
- Distribution
- Distribution to US states of CA, CO, FL, IL, IN, NY, MI, MN, TX, and WA, Poland and UK.
- Location
- Warsaw, IN
Frequently Asked Questions
What product was recalled? ▼
TRUMATCH CT CUT GUIDE KIT L- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation to assist in the positioning of a joint replacement component intra-operatively and in guiding the marking of bone before cutting Part Number: 420915. Recalled by DePuy Orthopaedics, Inc.. Units affected: 5 cases ( 1/cs).
Why was this product recalled? ▼
Software: A coding error associated with the Fast3D Segmentation software. During the scanning process the three images required of the hip, knee and ankle may not be aligned, causing the anatomic landmark locations to potentially result in limb malalignment.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on September 8, 2021. Severity: Moderate. Recall number: Z-2398-2021.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11