Severity
Moderate
FDA Devices recall · Reported September 8, 2021
Software: A coding error associated with the Fast3D Segmentation software. During the scanning process the three images required of the hip, knee and ankle may not be aligned, ca…
DePuy Orthopaedics, Inc. recalled TRUMATCH CT CUT GUIDE KIT L- TruMatch Patient Specific Instruments are intended to be use… - a moderate-severity action.
TRUMATCH CT CUT GUIDE KIT L- TruMatch Patient Specific Instruments are intended to be use… was recalled by DePuy Orthopaedics, Inc. in September 8, 2021. Reason: Software: A coding error associated with the Fast3D Segmentation software. During the scanning process the t…. Check the official notice for the remedy. Verify recall #Z-2398-2021 with the FDA Devices before acting.
The recall
DePuy Orthopaedics, Inc. issued this moderate-severity FDA Devices recall-Software: A coding error associated with the Fast3D Segmentation software. During the scanning process the t….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2398-2021 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2398-2021) was formally reported on September 8, 2021, with the manufacturer initiating the action on August 10, 2021. It is classified under Moderate severity (Class II), with a current status of Terminated. DePuy Orthopaedics, Inc. is listed as the recalling firm, operating out of Warsaw, IN. Federal records list the affected scope as 5 cases ( 1/cs).
The documented reason for this recall is: Software: A coding error associated with the Fast3D Segmentation software. During the scanning process the three images required of the hip, knee and ankle may not be aligned, causing the anatomic landmark locations to… Distribution data in the federal record shows the product reached: Distribution to US states of CA, CO, FL, IL, IN, NY, MI, MN, TX, and WA, Poland and UK.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
5 cases ( 1/cs)
Related Recalls
6
3 from same agency
TRUMATCH CT CUT GUIDE KIT L- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation to assist in the positioning of a joint replacement component intra-operatively and in guiding the marking of bone before cutting Part Number: 420915
Software: A coding error associated with the Fast3D Segmentation software. During the scanning process the three images required of the hip, knee and ankle may not be aligned, causing the anatomic landmark locations to potentially result in limb malalignment.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2398-2021 |
| Date reported | September 8, 2021 |
| Date initiated | August 10, 2021 |
| Recalling firm | DePuy Orthopaedics, Inc. |
| Firm location | Warsaw, IN |
| Affected scope | 5 cases ( 1/cs) |
| Distribution | Distribution to US states of CA, CO, FL, IL, IN, NY, MI, MN, TX, and WA, Poland and UK. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported September 8, 2021. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.