Severity
Moderate
FDA Devices recall · Reported September 26, 2012
Customers may be storing objects on the base and/or around the column of the surgical table. This prohibited practice and misuse known by the customers can damage the override sw…
Steris Corporation recalled A mobile, electro-hydraulically operated, fluoroscopic/C-Arm compatible general surgical … - a moderate-severity action.
A mobile, electro-hydraulically operated, fluoroscopic/C-Arm compatible general surgical … was recalled by Steris Corporation in September 26, 2012. Reason: Customers may be storing objects on the base and/or around the column of the surgical table. This prohibited…. Check the official notice for the remedy. Verify recall #Z-2402-2012 with the FDA Devices before acting.
The recall
Steris Corporation issued this moderate-severity FDA Devices recall-Customers may be storing objects on the base and/or around the column of the surgical table. This prohibited….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2402-2012 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2402-2012) was formally reported on September 26, 2012, with the manufacturer initiating the action on February 17, 2010. It is classified under Moderate severity (Class II), with a current status of Terminated. Steris Corporation is listed as the recalling firm, operating out of Mentor, OH. Federal records list the affected scope as 11,044 Units.
The documented reason for this recall is: Customers may be storing objects on the base and/or around the column of the surgical table. This prohibited practice and misuse known by the customers can damage the override switch assembly located at the top of the … Distribution data in the federal record shows the product reached: Worldwide Distribution - US Nationwide including District of Columbia and Puerto Rico and the following countries: American Samoa, Argentina, Australia, Bahrain, Bangladesh, Bermuda, Brazil, Canada, China, Colombia, Co…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
11,044 Units
Related Recalls
6
3 from same agency
A mobile, electro-hydraulically operated, fluoroscopic/C-Arm compatible general surgical table. Product Usage: Product provides a flexible, easy-to-use, articulated posturing of a patient for either general, I.A., or specialized surgical procedures. There are movable components and accessories intended for use during diagnostic examinations or surgical procedures to support and position a patient.
Customers may be storing objects on the base and/or around the column of the surgical table. This prohibited practice and misuse known by the customers can damage the override switch assembly located at the top of the column cover. Depending upon the nature and severity of the damage to the electronics of the table, it is possible that the table could move without prompting by an operator .
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2402-2012 |
| Date reported | September 26, 2012 |
| Date initiated | February 17, 2010 |
| Recalling firm | Steris Corporation |
| Firm location | Mentor, OH |
| Affected scope | 11,044 Units |
| Distribution | Worldwide Distribution - US Nationwide including District of Columbia and Puerto Rico and the following countries: American Samoa, Argentina, Australia, Bahrain, Bangladesh, Bermuda, Brazil, Canada, China, Colombia, Costa Rica, Dominican … |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported September 26, 2012. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.