PlainRecalls
FDA Devices Moderate Class II Ongoing

Constellation Procedure Pak

Reported: August 23, 2023 Initiated: June 28, 2023 #Z-2406-2023

Product Description

Constellation Procedure Pak

Reason for Recall

Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench.

Details

Recalling Firm
Alcon Research, LTD.
Units Affected
10,580 units
Distribution
Worldwide distribution - US Nationwide and the countries of Aland Islands, Albania, Algeria, Andorran, Argentina, Armenia, Aruba, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Bermuda, Botswana, Brazil, Brunei Daruss., Bulgaria, Burma, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cura¿ao, Cyprus, Czech Republic, Denmark, Dominican Rep., Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, French Guayana, Georgia, Germany, Ghana, Greece, Guadeloupe, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Isle of Man, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Luxembourg, Macau, Malaysia, Malta, Martinique, Mauritius, Mexico, Moldova, Mongolia, Montenegro, Morocco, Namibia, Netherlands, New Caledonia, New Zealand, Nicaragua, North Macedonia, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Fed., Saint Martin, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad, Tobago, Tunisia, Turkey, Ukraine, United Kingdom, Uruguay, USA, USVI, Utd. Arab Emir., Uzbekistan, Vietnam, and Yemen.
Location
Sinking Spring, PA

Frequently Asked Questions

What product was recalled?
Constellation Procedure Pak. Recalled by Alcon Research, LTD.. Units affected: 10,580 units.
Why was this product recalled?
Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench.
Which agency issued this recall?
This recall was issued by the FDA Devices on August 23, 2023. Severity: Moderate. Recall number: Z-2406-2023.