PlainRecalls
FDA Devices Moderate Class II Ongoing

Dxl 9000 Access Immunoassay Analyzer, Catalog #C11137.

Reported: July 31, 2024 Initiated: June 4, 2024 #Z-2414-2024

Product Description

Dxl 9000 Access Immunoassay Analyzer, Catalog #C11137.

Reason for Recall

The analyzers were built with nonconforming USB cables made with brass contacts instead of specified gold contacts. Brass contacts may corrode over time, leading to loss of internal instrument communication and delayed patient results.

Details

Recalling Firm
Beckman Coulter, Inc.
Units Affected
143 devices
Distribution
Worldwide - US Nationwide distribution in the states of FL, IN, KS, NH, PA, and TX. The countries of Australia, Austria, Belgium, Brazil, Croatia, Czechia, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Switzerland, Taiwan, Province of China, and United Kingdom of Great Britain and Northern Ireland.
Location
Chaska, MN

Frequently Asked Questions

What product was recalled?
Dxl 9000 Access Immunoassay Analyzer, Catalog #C11137.. Recalled by Beckman Coulter, Inc.. Units affected: 143 devices.
Why was this product recalled?
The analyzers were built with nonconforming USB cables made with brass contacts instead of specified gold contacts. Brass contacts may corrode over time, leading to loss of internal instrument communication and delayed patient results.
Which agency issued this recall?
This recall was issued by the FDA Devices on July 31, 2024. Severity: Moderate. Recall number: Z-2414-2024.