PlainRecalls
FDA Devices Moderate Class II Ongoing

Lokomat Pro, Catalog Numbers 30003 LokomatPro L6.2 120V AD ; 30004 LokomatPro L6.2 120V Kombi; 30005 LokomatPro L6.2 120V PE; 30683 LokomatPro L6.2 110V FD PE; 30684 LokomatPro L6.2 110V FD Kombi; 32001 LokomatPro L6.2 110V FD AD Product Usage: The intended use of the Lokomat is to support treadmill training to treat patients with walking disabilities caused by neurological, muscular or bone-related disorders. isokinetic testing and evaluation system

Reported: July 1, 2020 Initiated: June 8, 2020 #Z-2418-2020

Product Description

Lokomat Pro, Catalog Numbers 30003 LokomatPro L6.2 120V AD ; 30004 LokomatPro L6.2 120V Kombi; 30005 LokomatPro L6.2 120V PE; 30683 LokomatPro L6.2 110V FD PE; 30684 LokomatPro L6.2 110V FD Kombi; 32001 LokomatPro L6.2 110V FD AD Product Usage: The intended use of the Lokomat is to support treadmill training to treat patients with walking disabilities caused by neurological, muscular or bone-related disorders. isokinetic testing and evaluation system

Reason for Recall

The motor controller may fail, which can potentially lead to an error in functionality of the Body Weights Support Rope, which means that the rope can go up or down in an uncontrolled way.

Details

Recalling Firm
Hocoma AG
Units Affected
494
Distribution
Worldwide distribution - US Nationwide distribution including in the states of FL, GA, IL, MI, MN, MO, NE, NY, TX.
Location
Volketswil, N/A

Frequently Asked Questions

What product was recalled?
Lokomat Pro, Catalog Numbers 30003 LokomatPro L6.2 120V AD ; 30004 LokomatPro L6.2 120V Kombi; 30005 LokomatPro L6.2 120V PE; 30683 LokomatPro L6.2 110V FD PE; 30684 LokomatPro L6.2 110V FD Kombi; 32001 LokomatPro L6.2 110V FD AD Product Usage: The intended use of the Lokomat is to support treadmill training to treat patients with walking disabilities caused by neurological, muscular or bone-related disorders. isokinetic testing and evaluation system. Recalled by Hocoma AG. Units affected: 494.
Why was this product recalled?
The motor controller may fail, which can potentially lead to an error in functionality of the Body Weights Support Rope, which means that the rope can go up or down in an uncontrolled way.
Which agency issued this recall?
This recall was issued by the FDA Devices on July 1, 2020. Severity: Moderate. Recall number: Z-2418-2020.