Severity
Critical
FDA Devices recall · Reported August 14, 2024
CADD-Solis Ambulatory Infusion Pumps, with software versions before v4.3, may have the following issues:1. Upstream Occlusion, 2. Stop and Power Keys Unresponsive, 3. Manual Mode …
Smiths Medical Asd Inc. recalled CADD-Solis Ambulatory Infusion Pumps, Model 2100. Intended for the following uses: Indic… - a critical-severity action.
CADD-Solis Ambulatory Infusion Pumps, Model 2100. Intended for the following uses: Indic… was recalled by Smiths Medical Asd Inc. in August 14, 2024. Reason: CADD-Solis Ambulatory Infusion Pumps, with software versions before v4.3, may have the following issues:1. Up…. Check the official notice for the remedy. Verify recall #Z-2419-2024 with the FDA Devices before acting.
The recall
Smiths Medical Asd Inc. issued this critical-severity FDA Devices recall — CADD-Solis Ambulatory Infusion Pumps, with software versions before v4.3, may have the following issues:1. Up….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2419-2024 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2419-2024) was formally reported on August 14, 2024, with the manufacturer initiating the action on February 27, 2024. It is classified under Critical severity (Class I), with a current status of Ongoing. Smiths Medical Asd Inc. is listed as the recalling firm, operating out of Minneapolis, MN. Federal records list the affected scope as 9284 units.
The documented reason for this recall is: CADD-Solis Ambulatory Infusion Pumps, with software versions before v4.3, may have the following issues:1. Upstream Occlusion, 2. Stop and Power Keys Unresponsive, 3. Manual Mode Air Detector, 4. Single Bubble Air Detec… Distribution data in the federal record shows the product reached: Worldwide. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices — clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category — 40,409 medical devices recalls on record
Of 100,165 recalls in the database, 23,668 are high severity, 72,097 moderate, and 4,400 low. This recall is classified high severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Critical
Affected scope
9284 units
Related Recalls
6
6 from same agency
CADD-Solis Ambulatory Infusion Pumps, Model 2100. Intended for the following uses: Indicated for the following uses: Intra-arterial, subcutaneous, intraperitoneal, in close proximity to nerves, into an intraoperative site, epidural space or subarachnoid space. Therapies that require a continuous rate of infusion, intermittent bolus, and/or with patient-controlled demand doses.
CADD-Solis Ambulatory Infusion Pumps, with software versions before v4.3, may have the following issues:1. Upstream Occlusion, 2. Stop and Power Keys Unresponsive, 3. Manual Mode Air Detector, 4. Single Bubble Air Detection, 5. Error Codes Not Displayed at Power Up, 6. Audible Alarm, 7. Low Sensitivity Air in Line Detection Threshold, 8. PharmGuard Server Password. Smiths Medical corrected many of the issues included in this notification in previous software updates and the corrections were carried forward into all subsequent software releases. Please ensure you have the most recent CADD software (v4.3 - 2023) installed on your pumps.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Critical (Class I) |
| Status | Ongoing |
| Recall number | Z-2419-2024 |
| Date reported | August 14, 2024 |
| Date initiated | February 27, 2024 |
| Recalling firm | Smiths Medical Asd Inc. |
| Firm location | Minneapolis, MN |
| Affected scope | 9284 units |
| Distribution | Worldwide |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Keep tracking product safety across the federal recall archive.
Recall Checker
Search the full archive by product name, brand, or recall number across every agency.
Check a product →
FDA Devices recalls
Every recall issued by FDA Devices, newest first.
Browse the feed →
RecallRadar
Live feed of the latest recalls across the FDA, CPSC and NHTSA — filter by agency and severity.
View the live feed →
Rankings
The largest recalls by units affected and the most-recalled product categories.
See the rankings →
Browse by category
Find recalls by product type to spot recurring defect patterns.
All categories →
What to do next
A step-by-step guide to refunds, repairs, and returns after a recall.
Read the guide →
Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Medline Industries, LP · 2026-06-03
Abiomed, Inc. · 2026-06-03
GE Medical Systems Information Technologies Inc · 2026-06-03
Compare this recall with GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L… →
Source: FDA, CPSC, and NHTSA federal recall databases. This recall: FDA Devices, reported August 14, 2024.
Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of June 2026.