Nihon Kohden NKV-550 Ventilator, REF: NKV-550-U, CE 0344, UDI: (01) 00843685100012 The Nihon Kohden NKV-550 Series Ventilator System is intended to provide continuous ventilation for adult, pediatric and neonatal patients who require invasive or noninvasive respiratory support.
Reported: September 15, 2021 Initiated: July 20, 2021 #Z-2435-2021
Product Description
Nihon Kohden NKV-550 Ventilator, REF: NKV-550-U, CE 0344, UDI: (01) 00843685100012 The Nihon Kohden NKV-550 Series Ventilator System is intended to provide continuous ventilation for adult, pediatric and neonatal patients who require invasive or noninvasive respiratory support.
Reason for Recall
Potential defective gas inlet pressure regulator within ventilator which could result in gas releasing noise from the bottom of the ventilator, a High Air/O2 Inlet Pressure alarm, a High/Low O2 alarm, or could not pass device Check prior to use.
Details
- Recalling Firm
- NIHON KOHDEN ORANGEMED, INC
- Units Affected
- 1,999 units
- Distribution
- Worldwide Distribution: US (nationwide) to states of: AZ, CA, NY, MA, VT, LA, DC, PA, FL, UT and OUS (countries) of: Brazil, Japan, Ireland, Poland, Scotland, Germany, Kazakhstan, Spain, Slovenia, Germany/Bulgaria, Russia, India, Palestine, Libya, Jordan Kuwait, Lebanon, Syria, Turkey, South Africa, United Arab Emirates, Mexico, Colombia, Peru, Costa Rica, Barbados, Ecuador, Bahamas, Bolivia, St Kitts & Nevis, St. Martin, Chile, Thailand, Laos, Philippines, Malaysia, Nepal and United Kingdom.
- Location
- Santa Ana, CA
Frequently Asked Questions
What product was recalled? ▼
Nihon Kohden NKV-550 Ventilator, REF: NKV-550-U, CE 0344, UDI: (01) 00843685100012 The Nihon Kohden NKV-550 Series Ventilator System is intended to provide continuous ventilation for adult, pediatric and neonatal patients who require invasive or noninvasive respiratory support.. Recalled by NIHON KOHDEN ORANGEMED, INC. Units affected: 1,999 units.
Why was this product recalled? ▼
Potential defective gas inlet pressure regulator within ventilator which could result in gas releasing noise from the bottom of the ventilator, a High Air/O2 Inlet Pressure alarm, a High/Low O2 alarm, or could not pass device Check prior to use.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on September 15, 2021. Severity: Moderate. Recall number: Z-2435-2021.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11