Severity
Moderate
FDA Devices recall · Reported August 30, 2023
As part of the firm's ongoing quality monitoring (see internal complaint # OCC72574), Randox Laboratories have restandardized Ferritin in Liquid Protein Calibrators, IT2691, to r…
Randox Laboratories Ltd. recalled Randox Liquid Protein Calibrators-SP CAL-(LIQ)-In vitro diagnostic product used for the c… - a moderate-severity action.
Randox Liquid Protein Calibrators-SP CAL-(LIQ)-In vitro diagnostic product used for the c… was recalled by Randox Laboratories Ltd. in August 30, 2023. Reason: As part of the firm's ongoing quality monitoring (see internal complaint # OCC72574), Randox Laboratories ha…. Check the official notice for the remedy. Verify recall #Z-2448-2023 with the FDA Devices before acting.
The recall
Randox Laboratories Ltd. issued this moderate-severity FDA Devices recall-As part of the firm's ongoing quality monitoring (see internal complaint # OCC72574), Randox Laboratories ha….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2448-2023 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2448-2023) was formally reported on August 30, 2023, with the manufacturer initiating the action on July 19, 2023. It is classified under Moderate severity (Class II), with a current status of Ongoing. Randox Laboratories Ltd. is listed as the recalling firm, operating out of Crumlin (North), N/A. Federal records list the affected scope as 166 units.
The documented reason for this recall is: As part of the firm's ongoing quality monitoring (see internal complaint # OCC72574), Randox Laboratories have restandardized Ferritin in Liquid Protein Calibrators, IT2691, to reference material NISBC 19/118. The cali… Distribution data in the federal record shows the product reached: CA, FL, IL, ME, MI, NC, OH, PA,NJ,. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
166 units
Related Recalls
6
3 from same agency
Randox Liquid Protein Calibrators-SP CAL-(LIQ)-In vitro diagnostic product used for the calibration of ASO, Complement C3, Complement C4, CRP, Ferritin, Haptoglobin, IgM, Prealbumin, RF and Transferrin assays Catalogue Number: IT2691
As part of the firm's ongoing quality monitoring (see internal complaint # OCC72574), Randox Laboratories have restandardized Ferritin in Liquid Protein Calibrators, IT2691, to reference material NISBC 19/118. The calibrators lots 2112IT-2116IT, packed into batches 627222, 627224, 634886 and 634887 have been reassigned as part of the restandardization. Following this restandardization, Ferritin results for Quality Control material and patient samples recovered erroneously higher than the targeted calibrator values by approximately +10% across the assay range following.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-2448-2023 |
| Date reported | August 30, 2023 |
| Date initiated | July 19, 2023 |
| Recalling firm | Randox Laboratories Ltd. |
| Firm location | Crumlin (North), N/A |
| Affected scope | 166 units |
| Distribution | CA, FL, IL, ME, MI, NC, OH, PA,NJ, |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported August 30, 2023. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.