Severity
Moderate
FDA Devices recall · Reported September 2, 2015
Issue with the algorithm for ROI contraction in RayStation 4.7 when non-uniform contraction distances are used. The ROI contraction tool uses six distances as input: right/left, i…
Raysearch Laboratories Ab recalled RayStation 4.7, version 4.7.1, used with Structure Definition or Automatic Breast Plannin… - a moderate-severity action.
RayStation 4.7, version 4.7.1, used with Structure Definition or Automatic Breast Plannin… was recalled by Raysearch Laboratories Ab in September 2, 2015. Reason: Issue with the algorithm for ROI contraction in RayStation 4.7 when non-uniform contraction distances are use…. Check the official notice for the remedy. Verify recall #Z-2457-2015 with the FDA Devices before acting.
The recall
Raysearch Laboratories Ab issued this moderate-severity FDA Devices recall-Issue with the algorithm for ROI contraction in RayStation 4.7 when non-uniform contraction distances are use….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2457-2015 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2457-2015) was formally reported on September 2, 2015, with the manufacturer initiating the action on May 27, 2015. It is classified under Moderate severity (Class II), with a current status of Terminated. Raysearch Laboratories Ab is listed as the recalling firm, operating out of Stockholm, N/A. Federal records list the affected scope as 126 programs.
The documented reason for this recall is: Issue with the algorithm for ROI contraction in RayStation 4.7 when non-uniform contraction distances are used. The ROI contraction tool uses six distances as input: right/left, inferior/superior and posterior/anterior.… Distribution data in the federal record shows the product reached: Distributed in CO, IL, MI, NC, OH,TN, and WA.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
126 programs
Related Recalls
6
3 from same agency
RayStation 4.7, version 4.7.1, used with Structure Definition or Automatic Breast Planning. Radiation Therapy Treatment Planning System, for treatment planning and analysis of radiation therapy.
Issue with the algorithm for ROI contraction in RayStation 4.7 when non-uniform contraction distances are used. The ROI contraction tool uses six distances as input: right/left, inferior/superior and posterior/anterior. These contraction distances are, in error, pairwise reversed, i.e. right is interpreted as left, inferior is interpreted as superior, and posterior is interpreted as anterior.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2457-2015 |
| Date reported | September 2, 2015 |
| Date initiated | May 27, 2015 |
| Recalling firm | Raysearch Laboratories Ab |
| Firm location | Stockholm, N/A |
| Affected scope | 126 programs |
| Distribution | Distributed in CO, IL, MI, NC, OH,TN, and WA. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported September 2, 2015. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.