Wheel caster set screw for Ferno iNX Wheeled Stretcher. Ferno iNX Wheeled Stretcher for pre-hospital patient transport.

Reported: September 2, 2015 Initiated: May 1, 2015 #Z-2461-2015

Product Description

Wheel caster set screw for Ferno iNX Wheeled Stretcher. Ferno iNX Wheeled Stretcher for pre-hospital patient transport.

Reason for Recall

The firm discovered an improperly installed set screw in one caster arm assembly on a unit in production. The screw assists in preventing the bearing cup on the caster assembly from backing out of the caster arm during use.

Details

Recalling Firm
Ferno-Washington Inc
Units Affected
26 units
Distribution
Distributed in the states of AZ, CO, IA, MD, OH, PA & WA, and the countries of Australia, Canada & United Kingdom.
Location
Wilmington, OH

Frequently Asked Questions

What product was recalled?
Wheel caster set screw for Ferno iNX Wheeled Stretcher. Ferno iNX Wheeled Stretcher for pre-hospital patient transport.. Recalled by Ferno-Washington Inc. Units affected: 26 units.
Why was this product recalled?
The firm discovered an improperly installed set screw in one caster arm assembly on a unit in production. The screw assists in preventing the bearing cup on the caster assembly from backing out of the caster arm during use.