PlainRecalls
FDA Devices Moderate Class II Ongoing

BD Gravity IV Set 3-port Closed Manifold Back Check Valve 3 Needle-Free Y-sites, REF: CM42500E-07; BD Gravity IV Set 3-port Closed Stopock Manifold Back Check Valve 2 Needle-Free Y-sites, REF: CS42522E-07; BD GRAVITY 10 DP V/NV CHECK VALVE 3-WAY STOPCOCK 4 Smartsite VALVES & SmartSite EXT SET, REF: 42407E-05

Reported: September 22, 2021 Initiated: August 2, 2021 #Z-2463-2021

Product Description

BD Gravity IV Set 3-port Closed Manifold Back Check Valve 3 Needle-Free Y-sites, REF: CM42500E-07; BD Gravity IV Set 3-port Closed Stopock Manifold Back Check Valve 2 Needle-Free Y-sites, REF: CS42522E-07; BD GRAVITY 10 DP V/NV CHECK VALVE 3-WAY STOPCOCK 4 Smartsite VALVES & SmartSite EXT SET, REF: 42407E-05

Reason for Recall

Needle-free connector valves, either standalone or attached to extension sets, may have issues including difficulty to flush, flow issues, and partial or total occlusions, which could lead to therapy delay. Follow-up activates not required if product has already been used or was safely connected to the female luer of connecting device.

Details

Recalling Firm
CAREFUSION
Units Affected
24,360
Distribution
Worldwide distribution - US Nationwide distribution in the states of FL, PA, NC, OH, SC, IL, RI, MO, NY, TN, AL, VA, PR, CA, NE, NJ, MI, MD, NM, LA, KY, NV, ME, OK, WA, MA, WV, TX, AR, OR, KS, GA, VT, IN, WY, AZ, CO, MN, WI, UT, SD, MS, ID, MT, ND, HI, IA, DC, AK, NH, CT, DE, GU and the countries of CA, NZ, TW, AU, BE, VN, AR, MY, TH, KR, ID, PH, SG, BH, SA, HK, IN, CN, BR.
Location
San Diego, CA

Frequently Asked Questions

What product was recalled?
BD Gravity IV Set 3-port Closed Manifold Back Check Valve 3 Needle-Free Y-sites, REF: CM42500E-07; BD Gravity IV Set 3-port Closed Stopock Manifold Back Check Valve 2 Needle-Free Y-sites, REF: CS42522E-07; BD GRAVITY 10 DP V/NV CHECK VALVE 3-WAY STOPCOCK 4 Smartsite VALVES & SmartSite EXT SET, REF: 42407E-05. Recalled by CAREFUSION. Units affected: 24,360.
Why was this product recalled?
Needle-free connector valves, either standalone or attached to extension sets, may have issues including difficulty to flush, flow issues, and partial or total occlusions, which could lead to therapy delay. Follow-up activates not required if product has already been used or was safely connected to the female luer of connecting device.
Which agency issued this recall?
This recall was issued by the FDA Devices on September 22, 2021. Severity: Moderate. Recall number: Z-2463-2021.