FDA Devices recall · Reported October 3, 2012
Catheter Irrigation Set labeled in part: ***24K Arthroscopy Outflow/Suction Tubing Set*** This Arthroscopy Tubing Set is intended for use, in conjunction with the ConMed Linvatec 24k Pump, for pump-induced outflow and/or suction of fluid from the joint during arthroscopic procedures.
ConMed Linvatec recalled their 24k Arthroscopy Outflow/Suction Tubing Set (Product Number 24k100)because there is a possibility that the product may have a breach in the Tyvek sea…
Linvatec Corp. dba ConMed Linvatec recalled Catheter Irrigation Set labeled in part: ***24K Arthroscopy Outflow/Suction Tubing Set***… - a moderate-severity action.
Catheter Irrigation Set labeled in part: ***24K Arthroscopy Outflow/Suction Tubing Set***… was recalled by Linvatec Corp. dba ConMed Linvatec in October 3, 2012. Reason: ConMed Linvatec recalled their 24k Arthroscopy Outflow/Suction Tubing Set (Product Number 24k100)because ther…. Check the official notice for the remedy. Verify recall #Z-2469-2012 with the FDA Devices before acting.
The recall
Linvatec Corp. dba ConMed Linvatec issued this moderate-severity FDA Devices recall-ConMed Linvatec recalled their 24k Arthroscopy Outflow/Suction Tubing Set (Product Number 24k100)because ther….
- Moderate
- severity level
- 30K units
- affected scope
- Class II
- classification
- October 3, 2012
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-2469-2012 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
Catheter Irrigation Set labeled in part: ***24K Arthroscopy Outflow/Suction Tubing Set*** This Arthroscopy Tubing Set is intended for use, in conjunction with the ConMed Linvatec 24k Pump, for pump-induced outflow and/or suction of fluid from the joint during arthroscopic procedures.
Reason for recall
ConMed Linvatec recalled their 24k Arthroscopy Outflow/Suction Tubing Set (Product Number 24k100)because there is a possibility that the product may have a breach in the Tyvek seal that could potentially compromise the sterility of the contents.
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2469-2012 |
| Date reported | October 3, 2012 |
| Date initiated | July 1, 2011 |
| Recalling firm | Linvatec Corp. dba ConMed Linvatec |
| Firm location | Largo, FL |
| Affected scope | 30064 units |
| Distribution | Worldwide Distribution--USA (nationwide) including the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NV, NY, OH, OR, PA, SC, TN, TX, UT, WA ,WI and Hawaii., and the countries of Australia Ca… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-2469-2012) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls