PlainRecalls
FDA Devices Moderate Class II Ongoing

Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers, Product Models: 785-SR-ILP, 785-SR-ILC, 830-SR-ILP, 830-SR-ILC,785-ER-ILC.

Reported: September 17, 2025 Initiated: August 4, 2025 #Z-2478-2025

Product Description

Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers, Product Models: 785-SR-ILP, 785-SR-ILC, 830-SR-ILP, 830-SR-ILC,785-ER-ILC.

Reason for Recall

Two new potential laser safety failure modes have been identified in the product design. Overcurrent of laser diode, and a fail-unsafe mode. The failure modes have never been found to occur in these models, whether in the field or factory, but could potentially occur under excessive force or extreme, continuous vibration.

Details

Recalling Firm
WASATCH PHOTONICS
Units Affected
192
Distribution
U.S.
Location
Morrisville, NC

Frequently Asked Questions

What product was recalled?
Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers, Product Models: 785-SR-ILP, 785-SR-ILC, 830-SR-ILP, 830-SR-ILC,785-ER-ILC.. Recalled by WASATCH PHOTONICS. Units affected: 192.
Why was this product recalled?
Two new potential laser safety failure modes have been identified in the product design. Overcurrent of laser diode, and a fail-unsafe mode. The failure modes have never been found to occur in these models, whether in the field or factory, but could potentially occur under excessive force or extreme, continuous vibration.
Which agency issued this recall?
This recall was issued by the FDA Devices on September 17, 2025. Severity: Moderate. Recall number: Z-2478-2025.