Ingenuity Digital PET/CT System, Model 882446 with the following UDI's: (01)00884838059559(21)900004 (01)00884838059559(21)900005 (01)00884838059559(21)900006 (01)00884838059559(21)900007 (01)00884838059559(21)900008 (01)00884838059559(21)900009 (01)00884838059559(21)900010 (01)00884838059559(21)900011 (01)00884838059559(21)900012 (01)00884838059559(21)900013 (01)00884838059559(21)900014 (01)00884838059559(21)900015 (01)00884838059559(21)900016 (01)00884838059559(21)900017 (01)00
Product Description
Ingenuity Digital PET/CT System, Model 882446 with the following UDI's: (01)00884838059559(21)900004 (01)00884838059559(21)900005 (01)00884838059559(21)900006 (01)00884838059559(21)900007 (01)00884838059559(21)900008 (01)00884838059559(21)900009 (01)00884838059559(21)900010 (01)00884838059559(21)900011 (01)00884838059559(21)900012 (01)00884838059559(21)900013 (01)00884838059559(21)900014 (01)00884838059559(21)900015 (01)00884838059559(21)900016 (01)00884838059559(21)900017 (01)00884838059559(21)900018 (01)00884838059559(21)900019 (01)00884838059559(21)900020 (01)00884838059559(21)900021 (01)00884838059559(21)900022 (01)00884838059559(21)900023 (01)00884838059559(21)900024 (01)00884838059559(21)900025 (01)00884838059559(21)900026 (01)00884838059559(21)900027 (01)00884838059559(21)900028 (01)00884838059559(21)900029 (01)00884838059559(21)900030 (01)00884838059559(21)900031 (01)00884838059559(21)900032 (01)00884838059559(21)900033 (01)00884838059559(21)900034 (01)00884838059559(21)900035 Product Usage: The system is intended to image the whole body, heart, brain, lung, gastrointestinal, bone, lymphatic, and other major organs for a wide range of patient types, sizes, and extent of diseases. Both subsystems can also be operated as fully functional, independent diagnostic tools, including application of the CT scanner for diagnosis and for use in radiation therapy planning.
Reason for Recall
The block used to connect the patient pallet to the couch may be assembled incorrectly. A screw may be misaligned in the patient pallet.
Details
- Recalling Firm
- Philips Medical Systems (Cleveland) Inc
- Units Affected
- 32 units
- Distribution
- Worldwide Distribution - US Nationwide in the states of CA, GA, LA, MD, MI, NY, OH, TN, TX, VT, Washington DC & WI Internationally - Canada, Mexico, Belgium, France, Israel, Luxembourg, Netherlands, Singapore, Spain & Switzerland
- Location
- Cleveland, OH
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11