Severity
Moderate
FDA Devices recall · Reported July 15, 2020
A software issue was identified which could result in user or patient injury, or may adversely impact laser performance. It is possible that an unintended release of laser light (…
Ra Medical Systems, Inc. recalled DABRA Laser (RA-308 Excimer Laser). The DABRA Laser and DABRA Catheter is an excimer lase… - a moderate-severity action.
DABRA Laser (RA-308 Excimer Laser). The DABRA Laser and DABRA Catheter is an excimer lase… was recalled by Ra Medical Systems, Inc. in July 15, 2020. Reason: A software issue was identified which could result in user or patient injury, or may adversely impact laser p…. Check the official notice for the remedy. Verify recall #Z-2497-2020 with the FDA Devices before acting.
The recall
Ra Medical Systems, Inc. issued this moderate-severity FDA Devices recall-A software issue was identified which could result in user or patient injury, or may adversely impact laser p….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2497-2020 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2497-2020) was formally reported on July 15, 2020, with the manufacturer initiating the action on January 23, 2020. It is classified under Moderate severity (Class II), with a current status of Terminated. Ra Medical Systems, Inc. is listed as the recalling firm, operating out of Carlsbad, CA. Federal records list the affected scope as 79 units.
The documented reason for this recall is: A software issue was identified which could result in user or patient injury, or may adversely impact laser performance. It is possible that an unintended release of laser light (radiation) may occur, injuring the user … Distribution data in the federal record shows the product reached: US: AL, AZ, CA, FL, GA, ID, IL, IN, KS, KY, LA, MI, MD, NC, NV, NY, OK, OR, PA, TN,TX, WI. OUS: Italy and Japan.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,422 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
79 units
Related Recalls
6
0 from same agency
DABRA Laser (RA-308 Excimer Laser). The DABRA Laser and DABRA Catheter is an excimer laser ultraviolet light source and delivery system, collectively the DABRA Laser System". The system is designed to deliver the energy to the surgical site to ablate unwanted material. The optical conduit is used in conjunction with existing endovascular accessories and tools to complete a surgical procedure. The software is programmed onto an EEPROM chip, which is a component of the DABRA Excimer Laser RA-308 system. Catheter for crossing total occlusions in vascular system.
A software issue was identified which could result in user or patient injury, or may adversely impact laser performance. It is possible that an unintended release of laser light (radiation) may occur, injuring the user or patient.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2497-2020 |
| Date reported | July 15, 2020 |
| Date initiated | January 23, 2020 |
| Recalling firm | Ra Medical Systems, Inc. |
| Firm location | Carlsbad, CA |
| Affected scope | 79 units |
| Distribution | US: AL, AZ, CA, FL, GA, ID, IL, IN, KS, KY, LA, MI, MD, NC, NV, NY, OK, OR, PA, TN,TX, WI. OUS: Italy and Japan. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Keep tracking product safety across the federal recall archive.
Recall Checker
Search the full archive by product name, brand, or recall number across every agency.
Check a product →
FDA Devices recalls
Every recall issued by FDA Devices, newest first.
Browse the feed →
RecallRadar
Live feed of the latest recalls across the FDA, CPSC and NHTSA, filter by agency and severity.
View the live feed →
Rankings
The largest recalls by units affected and the most-recalled product categories.
See the rankings →
Browse by category
Find recalls by product type to spot recurring defect patterns.
All categories →
What to do next
A step-by-step guide to refunds, repairs, and returns after a recall.
Read the guide →
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-07-02
· 2026-07-02
· 2026-06-25
· 2026-06-25
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Source: FDA, CPSC, and NHTSA federal recall databases. This recall: FDA Devices, reported July 15, 2020.
Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.