Philips Allura System configured with a 1 Phase UPS. Allura family (R1.x, R2.x). Labeled as the following with corresponding model numbers: 1. Allura Xper FD10, Model Number 722003 2. Allura Xper FD10/10, Model Number 722005 3. Allura Xper FD20, Model Number 722006 4. Allura Xper FD20 Biplane, Model Number 722008 5. Allura Xper FD10, Model Number 722010 6. Allura Xper FD10/10, Model Number 722011 7. Allura Xper FD20, Model Number 722012 8. Allura Xper FD20 Biplane, M
Product Description
Philips Allura System configured with a 1 Phase UPS. Allura family (R1.x, R2.x). Labeled as the following with corresponding model numbers: 1. Allura Xper FD10, Model Number 722003 2. Allura Xper FD10/10, Model Number 722005 3. Allura Xper FD20, Model Number 722006 4. Allura Xper FD20 Biplane, Model Number 722008 5. Allura Xper FD10, Model Number 722010 6. Allura Xper FD10/10, Model Number 722011 7. Allura Xper FD20, Model Number 722012 8. Allura Xper FD20 Biplane, Model Number 722013 9. Allura Xper FD20 OR Table, Model Number 722023 10. Allura Xper FD10, Model Number 722026 11. Allura Xper FD10/10, Model Number 722027 12. Allura Xper FD20, Model Number 722028 13. Allura Xper FD20/10, Model Number 722029 14. Allura Xper FD20 OR Table, Model Number 722035 15. Allura Xper FD20/20, Model Number 722038 16. Allura Xper FD20/15, Model Number 722058 17. Allura Xper FD20/15 OR Table, Model Number 722059.
Reason for Recall
Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.
Details
- Recalling Firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Units Affected
- 1,006 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Korea, Republic of, Kosovo, Latvia, Lithuania, Macedonia, Maldives, Malta, Mexico, Moldova, Netherlands, New Caledonia, New Zealand, Nicaragua, Norway, Panama, Paraguay, Peru, Philippines, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tunisia, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Viet Nam.
- Location
- Best, N/A
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11