Severity
Moderate
FDA Devices recall · Reported August 17, 2016
Software defect, where either an Above Assay Range or an inaccurate value could potentially be reported, in the Stratus CS Acute Care Diagnostics System regarding TestPak calibrat…
Siemens Healthcare Diagnostics Inc recalled Stratus¿ CS STAT Fluorometric Analyzer-microprocessor-controlled instrument that measure… - a moderate-severity action.
Stratus¿ CS STAT Fluorometric Analyzer-microprocessor-controlled instrument that measure… was recalled by Siemens Healthcare Diagnostics Inc in August 17, 2016. Reason: Software defect, where either an Above Assay Range or an inaccurate value could potentially be reported, in t…. Check the official notice for the remedy. Verify recall #Z-2525-2016 with the FDA Devices before acting.
The recall
Siemens Healthcare Diagnostics Inc issued this moderate-severity FDA Devices recall-Software defect, where either an Above Assay Range or an inaccurate value could potentially be reported, in t….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2525-2016 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2525-2016) was formally reported on August 17, 2016, with the manufacturer initiating the action on June 22, 2016. It is classified under Moderate severity (Class II), with a current status of Terminated. Siemens Healthcare Diagnostics Inc is listed as the recalling firm, operating out of Norwood, MA. Federal records list the affected scope as 977 units.
The documented reason for this recall is: Software defect, where either an Above Assay Range or an inaccurate value could potentially be reported, in the Stratus CS Acute Care Diagnostics System regarding TestPak calibration and quality control (QC) status when… Distribution data in the federal record shows the product reached: Worldwide Distribution -- USA, U.A.E., Algeria, Armenia, Australia, Austria, Bangladesh, Belgium, Bosnia Herzeg., Botswana, Canada, Colombia, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, India, …. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
977 units
Related Recalls
6
3 from same agency
Stratus¿ CS STAT Fluorometric Analyzer-microprocessor-controlled instrument that measures certain analytes in body fluids for in vitro diagnostic use. Assays include ~hCG, CRP, CKMB, cTNI, D-Dimer, Myoglobin, and pBNP. SMN: 10444834, 10453531
Software defect, where either an Above Assay Range or an inaccurate value could potentially be reported, in the Stratus CS Acute Care Diagnostics System regarding TestPak calibration and quality control (QC) status when a TestPak is due to expire within 48 hours.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2525-2016 |
| Date reported | August 17, 2016 |
| Date initiated | June 22, 2016 |
| Recalling firm | Siemens Healthcare Diagnostics Inc |
| Firm location | Norwood, MA |
| Affected scope | 977 units |
| Distribution | Worldwide Distribution -- USA, U.A.E., Algeria, Armenia, Australia, Austria, Bangladesh, Belgium, Bosnia Herzeg., Botswana, Canada, Colombia, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Ireland, Italy, Ja… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported August 17, 2016. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.