Hamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator

Reported: September 27, 2023 Initiated: July 26, 2023 #Z-2529-2023

Product Description

Hamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator

Reason for Recall

Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.

Details

Recalling Firm
Hamilton Medical, Inc.
Units Affected
6 units
Distribution
Distribution US nationwide, including Puerto Rico. OUS: Not provided
Location
Reno, NV

Frequently Asked Questions

What product was recalled?
Hamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator. Recalled by Hamilton Medical, Inc.. Units affected: 6 units.
Why was this product recalled?
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.