Severity
Moderate
FDA Devices recall · Reported August 21, 2024
Beckman Coulter, Inc. has identified a rise in reagent carousel motion errors in Access 2 Immunoassay System Analyzers that appear to be caused by misaligned index wheels housed w…
Beckman Coulter, Inc. recalled Discrete photometric chemistry analyzer for clinical use. - a moderate-severity action.
Discrete photometric chemistry analyzer for clinical use. was recalled by Beckman Coulter, Inc. in August 21, 2024. Reason: Beckman Coulter, Inc. has identified a rise in reagent carousel motion errors in Access 2 Immunoassay System …. Check the official notice for the remedy. Verify recall #Z-2545-2024 with the FDA Devices before acting.
The recall
Beckman Coulter, Inc. issued this moderate-severity FDA Devices recall-Beckman Coulter, Inc. has identified a rise in reagent carousel motion errors in Access 2 Immunoassay System ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2545-2024 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2545-2024) was formally reported on August 21, 2024, with the manufacturer initiating the action on June 14, 2024. It is classified under Moderate severity (Class II), with a current status of Ongoing. Beckman Coulter, Inc. is listed as the recalling firm, operating out of Chaska, MN. Federal records list the affected scope as 59 units.
The documented reason for this recall is: Beckman Coulter, Inc. has identified a rise in reagent carousel motion errors in Access 2 Immunoassay System Analyzers that appear to be caused by misaligned index wheels housed within the reagent carousel module. Custo… Distribution data in the federal record shows the product reached: Worldwide - US Nationwide distribution in the states of NE, MT, LA, TX, NY, CA, MI, AL, IL the countries of Austria, Belgium, Georgia, India, Indonesia, Italy, Korea, Malaysia, Myanmar, Nigeria, Switzerland, Vietnam.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
59 units
Related Recalls
6
3 from same agency
Discrete photometric chemistry analyzer for clinical use.
Beckman Coulter, Inc. has identified a rise in reagent carousel motion errors in Access 2 Immunoassay System Analyzers that appear to be caused by misaligned index wheels housed within the reagent carousel module. Customers affected by this issue may observe an increase in reagent carousel motion errors or QSMC errors that occur while unloading or loading reagent packs during system initialization and sample processing, which causes a non-operational system status and cancels tests; and patient results may be delayed if these errors cancel test processing.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-2545-2024 |
| Date reported | August 21, 2024 |
| Date initiated | June 14, 2024 |
| Recalling firm | Beckman Coulter, Inc. |
| Firm location | Chaska, MN |
| Affected scope | 59 units |
| Distribution | Worldwide - US Nationwide distribution in the states of NE, MT, LA, TX, NY, CA, MI, AL, IL the countries of Austria, Belgium, Georgia, India, Indonesia, Italy, Korea, Malaysia, Myanmar, Nigeria, Switzerland, Vietnam. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported August 21, 2024. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.