OptiLite Multi-Use Holmium Laser Fibers, Supplied sterilized by ethylene oxide gas in peel-open packages Product Usage: Used with the Odyssey 30 Holmium Laser System for fragmentation of urinary calculi and soft tissue applications including incision/excision, ablation, and coagulation
Product Description
OptiLite Multi-Use Holmium Laser Fibers, Supplied sterilized by ethylene oxide gas in peel-open packages Product Usage: Used with the Odyssey 30 Holmium Laser System for fragmentation of urinary calculi and soft tissue applications including incision/excision, ablation, and coagulation
Reason for Recall
Cook Inc. is initiating a voluntary recall of the OptiLite Multi-Use Holmium Laser Fibers because it has been identified that the reprocessing instructions do not provide sufficiently detailed information for the cleaning, disinfection, and sterilization of these products.
Details
- Recalling Firm
- Cook Inc.
- Units Affected
- 10,415
- Distribution
- Worldwide Distribution - US Nationwide in the states of: AK AL AZ CA CO FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC NJ NY OH OR PA SC SD TN TX WA WI Foreign: Australia, Bolivia, Brazil, Canada, Chile, Columbia, Cyprus, Czech Republic, Germany, Dominican Republic, Greece, Hong Kong, Ireland, Israel, India, Iran, Jordan, Kuwait, Kazakhstan, Lebanon, Sri Lanka, Latvia, Libya, Mexico, Malaysia, New Zealand, Oman, Panama, Peru, Portugal, Serbia, Turkey, Trinidad and Tobago, Taiwan, and Venezuela. VA/DOD: Yes- see below
- Location
- Bloomington, IN
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11