PlainRecalls
FDA Devices Moderate Class II Completed

Brand Name: UNclog Eye Mask Product Name: UNclog Eye Mask Model/Catalog Number: AG23JW24 Serial # ag22 Product Description: Heated eye mask Component: Heating unit, wire

Reported: August 21, 2024 Initiated: February 8, 2024 #Z-2586-2024

Product Description

Brand Name: UNclog Eye Mask Product Name: UNclog Eye Mask Model/Catalog Number: AG23JW24 Serial # ag22 Product Description: Heated eye mask Component: Heating unit, wire

Reason for Recall

Due to a change in the safety clips for the heated eye masks, the wires to the safety clips may detach as a result of tugging and pulling and may cause a short in the heating element of the mask and could potentially cause burn injuries.

Details

Recalling Firm
EYE COMFORT CARE LLC
Units Affected
500
Distribution
U.S. Nationwide distribution in the states of AL, AZ, CA, CT, FL, KS, KY, MA, ME, MT, NC, ND, NJ, NM, NY, OH, OR, PA, SC, TX, VA, WA, and WI.
Location
Gales Creek, OR

Frequently Asked Questions

What product was recalled?
Brand Name: UNclog Eye Mask Product Name: UNclog Eye Mask Model/Catalog Number: AG23JW24 Serial # ag22 Product Description: Heated eye mask Component: Heating unit, wire. Recalled by EYE COMFORT CARE LLC. Units affected: 500.
Why was this product recalled?
Due to a change in the safety clips for the heated eye masks, the wires to the safety clips may detach as a result of tugging and pulling and may cause a short in the heating element of the mask and could potentially cause burn injuries.
Which agency issued this recall?
This recall was issued by the FDA Devices on August 21, 2024. Severity: Moderate. Recall number: Z-2586-2024.