Severity
Moderate
FDA Devices recall · Reported August 8, 2018
Several complaints were received for invalid calibration with low calibrator S1 while using VIDAS LH. Internal tests confirmed a decrease of the strips signal overtime for specif…
bioMerieux, Inc. recalled VIDAS LH, Reference Numbers 30406 and 30406-01. Poduct Usage: VIDAS LH is an automate… - a moderate-severity action.
VIDAS LH, Reference Numbers 30406 and 30406-01. Poduct Usage: VIDAS LH is an automate… was recalled by bioMerieux, Inc. in August 8, 2018. Reason: Several complaints were received for invalid calibration with low calibrator S1 while using VIDAS LH. Intern…. Check the official notice for the remedy. Verify recall #Z-2588-2018 with the FDA Devices before acting.
The recall
bioMerieux, Inc. issued this moderate-severity FDA Devices recall-Several complaints were received for invalid calibration with low calibrator S1 while using VIDAS LH. Intern….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2588-2018 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2588-2018) was formally reported on August 8, 2018, with the manufacturer initiating the action on March 1, 2018. It is classified under Moderate severity (Class II), with a current status of Terminated. bioMerieux, Inc. is listed as the recalling firm, operating out of Durham, NC. Federal records list the affected scope as 5,429 units.
The documented reason for this recall is: Several complaints were received for invalid calibration with low calibrator S1 while using VIDAS LH. Internal tests confirmed a decrease of the strips signal overtime for specific lots. Distribution data in the federal record shows the product reached: Worldwide Distribution - US Nationwide in the state of Arizona and North Carolina. Distributed internationally to Algeria, Algeria, Angola, Argentina, Bahrain, Bangladesh, Belarus, Bosnia-Herz., Brazil, Bulgaria, B…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
5,429 units
Related Recalls
6
3 from same agency
VIDAS LH, Reference Numbers 30406 and 30406-01. Poduct Usage: VIDAS LH is an automated quantitative test for use on the VIDAS family instruments for the determination of human luteinizing hormone in human serum or plasma (lithium heparin), using the ELFA technique (Enzyme Linked Fluorescent Assay)
Several complaints were received for invalid calibration with low calibrator S1 while using VIDAS LH. Internal tests confirmed a decrease of the strips signal overtime for specific lots.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2588-2018 |
| Date reported | August 8, 2018 |
| Date initiated | March 1, 2018 |
| Recalling firm | bioMerieux, Inc. |
| Firm location | Durham, NC |
| Affected scope | 5,429 units |
| Distribution | Worldwide Distribution - US Nationwide in the state of Arizona and North Carolina. Distributed internationally to Algeria, Algeria, Angola, Argentina, Bahrain, Bangladesh, Belarus, Bosnia-Herz., Brazil, Bulgaria, Burkina Faso, Cameroo… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported August 8, 2018. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.