Adult ThermoGard¿ Dual Dispersive Electrodes (for patients >15 kg.), No Cable, Catalog Number 51-7410. Intended to be used for the dispersion and return to the electrosurgical generator.
Reported: August 31, 2016 Initiated: May 18, 2016 #Z-2596-2016
Product Description
Adult ThermoGard¿ Dual Dispersive Electrodes (for patients >15 kg.), No Cable, Catalog Number 51-7410. Intended to be used for the dispersion and return to the electrosurgical generator.
Reason for Recall
A version of the dual dispersive electrodes may not be compatible with some electrosurgical generators, placing patients at risk for undetected pad lift and a potential burn.
Details
- Recalling Firm
- Conmed Corporation
- Units Affected
- 4,480 units
- Distribution
- Worldwide Distribution -- USA, including AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, & Puerto Rico; and, the countries of Australia, Canada, Belgium, Brunei Darussalam, China, France, Guatemala, Hong Kong, Israel, Italy, Japan, South Korea, Malaysia, Mexico, New Zealand, Norway, Portugal, Saudi Arabia, Spain, Sweden, and United Arab Emirates.
- Location
- Utica, NY
Frequently Asked Questions
What product was recalled? ▼
Adult ThermoGard¿ Dual Dispersive Electrodes (for patients >15 kg.), No Cable, Catalog Number 51-7410. Intended to be used for the dispersion and return to the electrosurgical generator.. Recalled by Conmed Corporation. Units affected: 4,480 units.
Why was this product recalled? ▼
A version of the dual dispersive electrodes may not be compatible with some electrosurgical generators, placing patients at risk for undetected pad lift and a potential burn.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on August 31, 2016. Severity: Moderate. Recall number: Z-2596-2016.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11