FDA Devices recall · Reported August 31, 2016

1) The HomeChoice automated peritoneal dialysis system. Dataplate located on the device is labeled in part: Baxter. 2) The HomeChoice PRO automated peritoneal dialysis system. Dataplate located on the device is labeled in part: Baxter. The HomeChoice/HomeChoice Pro Automated Personal Cycler peritoneal dialysis system is intended for automatic control of dialysate solution exchange in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.

In systems with version 10.4 software Initial Drain logic, the device will attempt to drain the patient to empty at the beginning of therapy (initial drain). The operator can stop…

Recall #
Z-2609-2016
Affected scope
*** 1) Product Codes 5C4471 and 5C4471R: Approximately 48,600 units; *** 2) Product Codes 5C8310 and 5C8310R: Approxiamtely 16,990 units
Initiated
October 7, 2013
Compiled from official public sources by the editorial team.
Verify with FDA Devices →
View my saved recalls

Baxter Healthcare Corp. recalled 1) The HomeChoice automated peritoneal dialysis system. Dataplate located on the device … - a moderate-severity action.

1) The HomeChoice automated peritoneal dialysis system. Dataplate located on the device … was recalled by Baxter Healthcare Corp. in August 31, 2016. Reason: In systems with version 10.4 software Initial Drain logic, the device will attempt to drain the patient to em…. Check the official notice for the remedy. Verify recall #Z-2609-2016 with the FDA Devices before acting.

The recall

Baxter Healthcare Corp. issued this moderate-severity FDA Devices recall-In systems with version 10.4 software Initial Drain logic, the device will attempt to drain the patient to em….

Moderate
severity level
Class II
classification
August 31, 2016
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-2609-2016 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This FDA Devices action (record #Z-2609-2016) was formally reported on August 31, 2016, with the manufacturer initiating the action on October 7, 2013. It is classified under Moderate severity (Class II), with a current status of Terminated. Baxter Healthcare Corp. is listed as the recalling firm, operating out of Deerfield, IL. Federal records list the affected scope as *** 1) Product Codes 5C4471 and 5C4471R: Approximately 48,600 units; *** 2) Product Codes 5C8310 and 5C8310R: Approxiamtely 16,990 units.

The documented reason for this recall is: In systems with version 10.4 software Initial Drain logic, the device will attempt to drain the patient to empty at the beginning of therapy (initial drain). The operator can stop, but cannot bypass, the active Initial … Distribution data in the federal record shows the product reached: Worldwide distribution. US nationwide including Puerto Rico; Australia, Canada, New Zealand, and Japan. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 40,322 medical devices recalls on record

-1,00001,0002,0003,0004,000 20052008201120142017202020232026 1,248
Severity2487872426High severity (most serious)Moderate severityLow severity
Where this recall sits in the database

Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Severity

Moderate

Affected scope

*** 1) Product Codes 5C4471 and 5C4471R: Approximately 48,600 units; *** 2) Product Codes 5C8310 and 5C8310R: Approxiamtely 16,990 units

Related Recalls

6

3 from same agency

Product description

1) The HomeChoice automated peritoneal dialysis system. Dataplate located on the device is labeled in part: Baxter. 2) The HomeChoice PRO automated peritoneal dialysis system. Dataplate located on the device is labeled in part: Baxter. The HomeChoice/HomeChoice Pro Automated Personal Cycler peritoneal dialysis system is intended for automatic control of dialysate solution exchange in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.

Reason for recall

In systems with version 10.4 software Initial Drain logic, the device will attempt to drain the patient to empty at the beginning of therapy (initial drain). The operator can stop, but cannot bypass, the active Initial Drain. This is to mitigate against Unintended Increased Intraperitoneal Volume (IIVP). This can cause serious problems in patients with unrelated a co-morbid condition of ascites.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-2609-2016
Date reported August 31, 2016
Date initiated October 7, 2013
Recalling firm Baxter Healthcare Corp.
Firm location Deerfield, IL
Affected scope *** 1) Product Codes 5C4471 and 5C4471R: Approximately 48,600 units; *** 2) Product Codes 5C8310 and 5C8310R: Approxiamtely 16,990 units
Distribution Worldwide distribution. US nationwide including Puerto Rico; Australia, Canada, New Zealand, and Japan

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-2609-2016) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
1) The HomeChoice automated peritoneal dialysis system. Dataplate located on the device is labeled in part: Baxter. 2) The HomeChoice PRO automated peritoneal dialysis system. Dataplate located on the device is labeled in part: Baxter. The HomeChoice/HomeChoice Pro Automated Personal Cycler peritoneal dialysis system is intended for automatic control of dialysate solution exchange in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.. Recalled by Baxter Healthcare Corp.. Units affected: *** 1) Product Codes 5C4471 and 5C4471R: Approximately 48,600 units; *** 2) Product Codes 5C8310 and 5C8310R: Approxiamtely 16,990 units.
Why was this product recalled?
In systems with version 10.4 software Initial Drain logic, the device will attempt to drain the patient to empty at the beginning of therapy (initial drain). The operator can stop, but cannot bypass, the active Initial Drain. This is to mitigate against Unintended Increased Intraperitoneal Volume (IIVP). This can cause serious problems in patients with unrelated a co-morbid condition of ascites.
Which agency issued this recall?
This recall was issued by the FDA Devices on August 31, 2016. Severity: Moderate. Recall number: Z-2609-2016.
Where was the recalled product distributed?
Distribution: Worldwide distribution. US nationwide including Puerto Rico; Australia, Canada, New Zealand, and Japan.
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-2609-2016) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nearby Recalls in This Category

Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →

Browse all Medical Devices recalls →

Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported August 31, 2016. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.