PlainRecalls
FDA Devices Critical Class I Ongoing

Reprocessed Inquiry Steerable Diagnostics Catheter for electrogram recording and cardiac stimulation during diagnostic electrophysiologic studies; Medline Item Numbers 81102RH 81104RH 81107RH 81174RH 81402RH 81404RH 81405RH 81472RH 81473RH 81474RH 81531RH 81532RH

Reported: October 8, 2025 Initiated: July 31, 2025 #Z-2612-2025

Product Description

Reprocessed Inquiry Steerable Diagnostics Catheter for electrogram recording and cardiac stimulation during diagnostic electrophysiologic studies; Medline Item Numbers 81102RH 81104RH 81107RH 81174RH 81402RH 81404RH 81405RH 81472RH 81473RH 81474RH 81531RH 81532RH

Reason for Recall

Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

Details

Recalling Firm
Medline Industries, LP
Units Affected
7913 eaches
Distribution
Domestic US distribution nationwide. International distribution pending. No international distribution.
Location
Northfield, IL

Frequently Asked Questions

What product was recalled?
Reprocessed Inquiry Steerable Diagnostics Catheter for electrogram recording and cardiac stimulation during diagnostic electrophysiologic studies; Medline Item Numbers 81102RH 81104RH 81107RH 81174RH 81402RH 81404RH 81405RH 81472RH 81473RH 81474RH 81531RH 81532RH. Recalled by Medline Industries, LP. Units affected: 7913 eaches.
Why was this product recalled?
Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.
Which agency issued this recall?
This recall was issued by the FDA Devices on October 8, 2025. Severity: Critical. Recall number: Z-2612-2025.